The development of a wearable pulse oximeter sensor and study of the calibration method / 生物医学工程学杂志
Artigo em Chinês | WPRIM (Pacífico Ocidental) | ID: wpr-294582
Biblioteca responsável: WPRO
ABSTRACT
The paper first analyses the principles of measurement of the two-wave oximeter and their limitations in technology. We propose to filter off motion interference from pulse oximeter signal using an algorithm based on the Beer-Lambert law that requires a three-wave probe (660 nm, 850 nm, and 940 nm). Based on the new algorithm, this paper describes the design principle of the circuitry and the software flowchart. Also, we study the calibration method of the pulse oximeter sensor and discuss the results in this paper.
Assuntos
Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Oxigênio / Padrões de Referência / Calibragem / Oxiemoglobinas / Oximetria / Desenho de Equipamento / Metabolismo Tipo de estudo: Estudo prognóstico Limite: Humanos Idioma: Chinês Revista: Journal of Biomedical Engineering Ano de publicação: 2009 Tipo de documento: Artigo
Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Oxigênio / Padrões de Referência / Calibragem / Oxiemoglobinas / Oximetria / Desenho de Equipamento / Metabolismo Tipo de estudo: Estudo prognóstico Limite: Humanos Idioma: Chinês Revista: Journal of Biomedical Engineering Ano de publicação: 2009 Tipo de documento: Artigo
...