Your browser doesn't support javascript.
loading
Fabrication and Optimization of a Fiber-optic Dosimeter for Proton Beam Therapy Dosimetry / 의학물리
Artigo em Coreano | WPRIM (Pacífico Ocidental) | ID: wpr-30107
Biblioteca responsável: WPRO
ABSTRACT
In this study, we have fabricated a fiber-optic dosimeter for a proton beam therapy dosimetry. We have measured scintillating lights with the various kinds of organic scintillators and selected the BCF-12 as a sensor-tip material due to its highest light output and peak/plateau ratio. To determine the optimum diameter of BCF-12, we have measured scintillating lights according to the energy losses of proton beams in a water phantom. Also, we determined the adequate length of organic scintillator by measuring scintillating lights according to the incident angles of proton beam. Using an optimized fiber-optic dosimeter, we have measured scintillating lights according to the dose rates and monitor units of proton accelerator.
Assuntos

Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Compostos Organotiofosforados / Prótons / Água / Fibras Ópticas / Terapia com Prótons / Luz Idioma: Coreano Revista: Korean Journal of Medical Physics Ano de publicação: 2010 Tipo de documento: Artigo
Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Compostos Organotiofosforados / Prótons / Água / Fibras Ópticas / Terapia com Prótons / Luz Idioma: Coreano Revista: Korean Journal of Medical Physics Ano de publicação: 2010 Tipo de documento: Artigo
...