Application value of computed radiography energy subtraction in detecting pulmonary soft-tissue:comparison with direct digital radiography / 中华放射医学与防护杂志
Article em Zh | WPRIM | ID: wpr-387783
Biblioteca responsável: WPRO
ABSTRACT
Objective To compare the efficiency of computed radiography energy subtraction (ES-CR) with direct digital radiography (DR) in detecting the pulmonary soft-tissue signals.Methods The adult chest phantom including TRG was imaged by ES-CR and DR separately and was estimated by 6 radiologists with 5-value-differentiation method. ROC was used to analyze the performance of two systems. Images of DR and ES-CR of 28 patients with small lung cancer were diagnosed by 3 radiologists. Results The area under the ROC curves was 0. 727 in ES-CR and 0.827 in DR (Z= 2.96,P < 0. 05). 28 small lung cancer patients were significantly different between two systems (x2 = 5.14,P <0. 05). Conclusions The true positive proportion of DR might be higher than that of ES-CR. ES-CR is inferior to DR in detecting pulmonary soft-tissue signals.