Commissioning of Mobetron mobile intra-operattve radiotherapy accelerator / 中华放射肿瘤学杂志
Artigo em Chinês | WPRIM (Pacífico Ocidental) | ID: wpr-394065
Biblioteca responsável: WPRO
ABSTRACT
Objective To commission a Mobetron intra-operative mobile accelerator and analyze the characteristics of its electron beams. Methods The dosimetrie characteristics of the electron beams genera-ted by Mobetron accelerator were measured and compared with those generated by conventional accelerator (Primus, Siemens). M oberton accelerator can generate electron beams of nominal energies of 4,6,9 and 12 MeV. The measurement items were as followings percentage depth dose perpendicular to water phantom sur-face and beam profiles parallel to water phantom surface, output factors, applicator leakage, electron beam at-tenuation made by lead blocks,and machine output calibration. The measurement devices included a three-dimensional ( 3 D) water scanning phantom, an electrometer, a 0.6 cm3 Farmer ionization chamber, a parallel-plate ionization chamber and solid water slabs. During measurement, all applicators of different tilt angles and diameters were attached to the machine head,and their ends were adjusted to be tangent to the phantom surface. Results Except for the 12 MeV,skin dose for all energies was no more than 90%. The skin dose was higher for Mobetron accelerator electron beams than for regular electron beams. The Dmax depth in water for a 10 cm flat applicator were 0.7,1.3,2.0 and 2.2 cm for the 4 energies,respectively. The depths of 90% dose were 1.0,1.8,2.7 and 3.6 cm, respectively. The selected flat applicator was just 1 cm larger than the tumor bed. But for the beveled applicators,the field flatness and symmetry became worse,and con-sequently,the applicator size had to be selected based on the isodose distribution. The leakage dose at 1 cm outside the applicator was 1.2% ,5.1%, 10.0% and 9.1%, respectively. The lead thickness for full block was 1.5,3.0,4.5 and 6.0 mm,respectively. Conclusions Through the commissioning of Mobetron accel-erator, the machine characteristics are understood, and the data for clinical implementation and routine quality assurance are acquired.

Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Radiation Oncology Ano de publicação: 2009 Tipo de documento: Artigo
Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Radiation Oncology Ano de publicação: 2009 Tipo de documento: Artigo
...