Performance verification of precision and accuracy of Olympus AU5421 detection system / 国际检验医学杂志
Artigo em Chinês | WPRIM (Pacífico Ocidental) | ID: wpr-457633
Biblioteca responsável: WPRO
ABSTRACT
Objective To analyze the precision and accuracy of the Olympus AU5421 automatic biochemical analyzer in order to verify the performance of this detection system declared by the manufacturer.Methods The Precision and Accuracy of User Au-thentication-Guide for Approval Second Edition(CLSI EP15-A2)was used to perform the routine detection on the Olympus AU5421 automatic biochemical analyzer.The systematic precision and accuracy were analyzed.Results Under this experiment condition,the precision and accuracy of the Olympus AU5421 automatic biochemical analyzer was consistent with the performance declared by the manufacturer.Conclusion The Olympus AU5421 automatic biochemical analyzer system has the high precision and good accuracy, and can be better applied in the clinical routine detection.

Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Tipo de estudo: Estudo diagnóstico Idioma: Chinês Revista: International Journal of Laboratory Medicine Ano de publicação: 2014 Tipo de documento: Artigo