Quality control research of active spot scanning proton and heavy ion accelerator's radiation field uniformity / 中华放射医学与防护杂志
Artigo em Chinês | WPRIM (Pacífico Ocidental) | ID: wpr-496869
Biblioteca responsável: WPRO
ABSTRACT
Objective To develop the method for testing the consistency of irradiation field produced by the active spot scanning proton and heavy ion accelerator.Methods Calibration of the EBT3 films were carried out with the calibrated ion beam to establish the dose calibration curve.According to the different proton and carbon ion energies (proton94.29,150.68,212.62 McV;carbon ion175.99,283.43,412.54 MeV/u),EBT3 films were located in the solid water phantoms in each therapy room,respectively.Finally,the irradiated EBT3 films were scanned and the radiation field size's deviation and flatness were analyzed.Results In different conditions,radiation field size's deviations were all less than 2 mm and the flatness parameters were all controlled below the 5%.Conclusions EBT3 films can be used to test the active spot scanning proton and heavy ion accelerator's radiation field uniformity.

Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Radiological Medicine and Protection Ano de publicação: 2016 Tipo de documento: Artigo