Study on the Adsorption of Disposable Transfusion Connective Tube and Infusion Needle for Single Use only to Nitroglycerin / 中国药房
China Pharmacy ; (12)2001.
Artigo em Chinês | WPRIM (Pacífico Ocidental) | ID: wpr-524527
Biblioteca responsável: WPRO
ABSTRACT

OBJECTIVE:

To study the adsorption of disposable transfusion connective tube and infusion needle for single use only to nitroglycerin in intravenous injection via minipump.

METHODS:

The intravenous injection via minipump was im-itated;the concentration change of nitroglycerin during the injection process was determined by the high performance liquid chromatography(HPLC)and ultraviolet spectrophotometry.

RESULTS:

Disposable transfusion connective tube and infusion needle for single use only had significant adsorption to nitroglycerin with an mean adsorption rate at(73.88?2.05)%within8hours.

CONCLUSION:

It is unsuitable to use those disposable transfusion connective tubes and infusion needles for single use only that have strong adsorption to nitroglycerin in the intravenous injection of nitroglycerin via minipump.

Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: China Pharmacy Ano de publicação: 2001 Tipo de documento: Artigo