Significance of Hemoglobin Detection Test Kit Using in Hospital Supply Center / 中华医院感染学杂志
Chinese Journal of Nosocomiology
; (24)2006.
Article
em Zh
| WPRIM
| ID: wpr-594663
Biblioteca responsável:
WPRO
ABSTRACT
OBJECTIVE To compare the difference of testing cleaning effectiveness in the operation instrument between the eye measurement and hemoglobin enzyme test.METHODS Five batches of instrument after cleaning were tested by two methods.RESULTS By eye inspection,all tested samples were qualified.However there were only 76 percent of samples qualified according to enzyme test.CONCLUSIONS Compared with eye inspection,enzyme test is a more sensitive method.Enzyme test can better assess the cleaning effectiveness than others.
Texto completo:
1
Base de dados:
WPRIM
Tipo de estudo:
Diagnostic_studies
Idioma:
Zh
Revista:
Chinese Journal of Nosocomiology
Ano de publicação:
2006
Tipo de documento:
Article