Performance Evaluation of BC-5500 Hematology Analyzer / 医疗卫生装备
Chinese Medical Equipment Journal
; (6)1993.
Artigo
em Chinês
| WPRIM (Pacífico Ocidental)
| ID: wpr-594997
Biblioteca responsável:
WPRO
ABSTRACT
0.99). DC reproducibility was good for neutrophils, lymphocytes monocytes, eosinophils and basophils. Comparison of the results by instrument with manual for normal samples in morphology, the correlation was better for neutrophils, lymphocytes and eosinophils (r0.968~0.983) ,good for monocytes(r=0.917), not good for basophils(r=0.659);The WBC scattergram would change and alarm flags would display when there are neutrophilic stab granulocytes, abnormal or atypical lymphocytes and immature cells in higher proportion.Conclusions The analyzer can be used to test blood cell parameters accurately and reliably. Its main performance indices accorded with the experimental requirements; The results were credible. It is necessary to check with microscopy for DC before reported when it were doubtted.
Texto completo:
Disponível
Base de dados:
WPRIM (Pacífico Ocidental)
Idioma:
Chinês
Revista:
Chinese Medical Equipment Journal
Ano de publicação:
1993
Tipo de documento:
Artigo