Dry heat sterilization of ophthalmic instruments using a commercially available convection broiler
Philippine Journal of Ophthalmology
; : 131-139, 2003.
Article
em En
| WPRIM
| ID: wpr-632283
Biblioteca responsável:
WPRO
ABSTRACT
This study is proposed as an alternative method of sterilization in developing countries Purpose: To determine the efficacy of using a commercially available convection broiler in sterilization of ophthalmic surgical instruments Methods: Surgical ophthalmic instruments were contaminated artificially with Staphylococcus epidermidis, Pseudomonas aeruginosa, and Aspergillus. The instruments were subjected to heat sterilization using the convection broiler at different temperature settings. After which, the instruments were cultured with appropriate media to determine contamination Results: 15 minutes heat sterilization at 260 degrees centigrade was enough to sterilize the contaminated ophthalmic surgical instruments Conclusion: Heat sterilization using a convection broiler provide a cheaper yet effective method of sterilization. (Author)
Texto completo:
1
Base de dados:
WPRIM
Limite:
Humans
Idioma:
En
Revista:
Philippine Journal of Ophthalmology
Ano de publicação:
2003
Tipo de documento: