Your browser doesn't support javascript.
loading
Eye model for inspecting the functional vision of IOLs / 国际眼科杂志(Guoji Yanke Zazhi)
International Eye Science ; (12): 1233-1236, 2006.
Article em Zh | WPRIM | ID: wpr-641717
Biblioteca responsável: WPRO
ABSTRACT
· AIM: To inspect and compare the functional vision of an aspheric intraocular lens (Tecnis) with those of conventional monofocal silicone and acrylic intraocular lens and multifocal intraocular lens (Array).· METHODS: The IOLs were tested in the eye model, which was designed to be optically equivalent to the theoretical eye model. The eye model is a combination of a spherical photographic lens with 35mm focal length ( IOL put in a water cell)and a charge coupled device (CCD) camera. The images constructed by the lenses are observed on a monitor of personal computer and the contrasts of the images are analyzed by using commercial image processing software.· RESULTS: The modulation transfer function of the eye model equals the scale produced by the theoretical eye model. The images constructed by changing the diameter of aperture stop and IOL.· CONCLUSION: The proposed eye model is useful for testing functional vision and for inspecting differences of intraocular lens.
Palavras-chave
Texto completo: 1 Base de dados: WPRIM Idioma: Zh Revista: International Eye Science Ano de publicação: 2006 Tipo de documento: Article
Texto completo: 1 Base de dados: WPRIM Idioma: Zh Revista: International Eye Science Ano de publicação: 2006 Tipo de documento: Article