Determination of serum progesterone by isotope dilution gas chromatography mass spectrometry / 中华检验医学杂志
Chinese Journal of Laboratory Medicine
; (12)2001.
Article
em Zh
| WPRIM
| ID: wpr-685325
Biblioteca responsável:
WPRO
ABSTRACT
Objective To develop a candidate reference method for the measurement of progesterone in human serum.Methods The serum sample is mixed with the internal standard [3,4-~(13)C_2] progesterone.After extraction with n-hexane and purified by a aqueous solution of 2-Hydroxypropyl-?- cyclodextrin (HP-?-CD),the serum progesterone and labeled progesterone are converted to the 3-enol heptafluorobutyrate and analyzed by gas chromatography mass spectrometry (GC/MS) with selected ion monitoring.The concentration of serum progesterone is calculated by bracketing method.Results The results gave coefficients of variation (CVs) of 0.69% to 2.12%.The analytical recoveries ranged from 98.3% to 100.1%.The results of measuring certified reference materials of serum progesterone are agree with the target value.Conclusion The procedure for measuring progesterone in serum is a highly accurate and precise method and may be used as a candidate reference method for serum progesterone assays.
Texto completo:
1
Base de dados:
WPRIM
Idioma:
Zh
Revista:
Chinese Journal of Laboratory Medicine
Ano de publicação:
2001
Tipo de documento:
Article