ABSTRACT
OBJECTIVE@#In order to solve alarm fatigue, the algorithm optimization strategies were researched to reduce false and worthless alarms.@*METHODS@#A four-lead arrhythmia analysis algorithm, a multiparameter fusion analysis algorithm, an intelligent threshold reminder, a refractory period delay technique were proposed and tested with collected 28 679 alarms in multi-center study.@*RESULTS@#The sampling survey indicate that the 80.8% of arrhythmia false alarms were reduced by the four-lead analysis, the 55.9% of arrhythmia and pulse false alarms were reduced by the multi-parameter fusion analysis, the 28.0% and 29.8% of clinical worthless alarms were reduced by the intelligent threshold and refractory period delay techniques respectively. Finally, the total quantity of alarms decreased to 12 724.@*CONCLUSIONS@#To increase the dimensionality of parametric analysis and control the alarm limits and delay time are conducive to reduce alarm fatigue in intensive care units.
Texto completo:
Disponível
Contexto em Saúde:
Agenda de Saúde Sustentável para as Américas
/
ODS3 - Saúde e Bem-Estar
Problema de saúde:
Objetivo 9: Redução de doenças não transmissíveis
/
Meta 3.4: Reduzir as mortes prematuras devido doenças não transmissíveis
Base de dados:
WPRIM (Pacífico Ocidental)
Assunto principal:
Arritmias Cardíacas
/
Alarmes Clínicos
/
Fadiga de Alarmes do Pessoal de Saúde
/
Unidades de Terapia Intensiva
/
Monitorização Fisiológica
Tipo de estudo:
Ensaio clínico controlado
Limite:
Humanos
Idioma:
Chinês
Revista:
Chinese Journal of Medical Instrumentation
Ano de publicação:
2020
Tipo de documento:
Artigo