Comparison of Aberrations in Korean Normal Eyes Measured With Two Different Aberrometers
Artigo em Coreano | WPRIM (Pacífico Ocidental) | ID: wpr-96516
Biblioteca responsável: WPRO
ABSTRACT

PURPOSE:

To compare ocular higher order aberrations measured by two different aberrometers in a sample of normal eyes.

METHODS:

We included 41 normal eyes of Koreans in this study. Ocular aberration data were obtained through three measurements per eye using Zywave and OPD-Scan devices. Spherical equivalent and higher order aberrations calculated in the central 6 mm zone and expressed as root mean square (RMS) values were analyzed.

RESULTS:

A comparison of measurements between the Zywave and OPD-Scan devices demonstrated no statistically significant differences in the RMS values of total higher order aberration (p=0.11), but significant differences were detected in the RMS values of total spherical aberration, total coma and total trefoil (p<0.01).

CONCLUSIONS:

The two different aberrometers that we tested are suitable for taking repeated measurements and are internally consistent, but not interchangeable.
Assuntos

Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Coma / Olho / Lotus Idioma: Coreano Revista: Journal of the Korean Ophthalmological Society Ano de publicação: 2009 Tipo de documento: Artigo