Your browser doesn't support javascript.
loading
Analysis of quality control test results of medical electron linear accelerators in Guangxi Province 2017—2019 / 中国辐射卫生
Artigo em Chinês | WPRIM (Pacífico Ocidental) | ID: wpr-974568
Biblioteca responsável: WPRO
ABSTRACT
Objective To understand the level of quality control in medical linear accelerator in Guangxi, and to provide technical support for the health administrative department to make related standards. Methods To select medical linear accelerators in Guangxi by random sampling method. Results A total of 118 accelerators were checked in 2017—2019, the initial inspection pass rates of medical linear accelerators were 56.8% and re-inspect pass rate were 96.6%. Conclusion The quality control level of the radiology institutions needs to be improved. The health administrative department should strengthen routine management,it is necessary to implement quality control in medical linear accelerator.

Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Radiological Health Ano de publicação: 2021 Tipo de documento: Artigo
Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Radiological Health Ano de publicação: 2021 Tipo de documento: Artigo
...