Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
Chinese Journal of Medical Instrumentation ; (6): 23-25, 2008.
Article in Chinese | WPRIM | ID: wpr-323230

ABSTRACT

This paper proposes a nonlinear method of detecting the blood oxygen saturation (SpO2) in a wide range on the basis of spectrophotometry, solves the nonlinear problem in detecting SpO2 in a wide range, and improves the applicable range and detection precision. CAN bus communication and DSP fast digital signal processing are utilized in the design of a non-invasive pulse blood oxygen saturation (SpO2) monitor in a wide range, which has a measure error below 3% within a range of 35% approximately 100%, and is applicable for clinical monitoring and detection of cosmonauts' physiological signals. This oximeter's components and their functions are also introduced in detail in the paper.


Subject(s)
Equipment Design , Oximetry , Methods
SELECTION OF CITATIONS
SEARCH DETAIL