ABSTRACT
Objective@#To establish the method for determination of tungsten and insoluble compounds in the air of workplace by inductively coupled plasma optical emission spectrometry (ICP-OES) .@*Methods@#The tungsten and insoluble compounds were digested by microwave digestion apparatus using nitric acid and hydrofluoric acid, detected by ICP-OES.@*Results@#The linearity of tungsten and insoluble compounds (as tungsten) were good at the range of 0.16-100.0 μg/ml, the minimum quantitation concentration was 0.11 mg/m3, the recovery was ranged from 81.9%-97.1%, the RSD of intra-and inter-batch precision were 0.5%-2.5% and 1.2%-4.4%, respectively.@*Conclusion@#The determination method meet the requirement of analysis and apply to the determination of tungsten and insoluble compounds in the air of workplace.
ABSTRACT
Objective@#To establish the method for determination of barium sulfate in the air of workplace. @*Methods@#The barium sulfate was collected by dichloride ethylene filter membrane and then processed by alkali fusion method. Inductively coupled plasma optical emission spectrometry (ICP-OES) was used for the detection of barium sulfate. @*Results@#The sampling efficiency was 100%, the linearity of ICP-OES was good at the range of 0.1~100.0 μg/mL, the recovery was ranged from 93.0%~97.8%, the RSD of intra- and inter-batch precision were 3.7%~7.6% and 4.7%~8.8%, respectively. @*Conclusion@#The sampling method and determination method meet the requirements of analysis and apply to the collection and determination of barium sulfate in the air of workplace.
ABSTRACT
Objective To establish a method for the determination of 8 heavy metal elements of Pb,Cd,As,Hg,Co,V,Se,Mo in iron dextran by inductively coupled plasma optical emission spectrometry (ICP-OES).Method Through selection of detection wavelengths,optimization of instrument parameters,verification of matrix effect interference and investigation of internal standard for correction of matrix effect interference,the 8 heavy metal elements were analyzed by ICP-OES.Results The recovery rate of the detected elements were increased from 83%-90% to 94%-100%,after correction of matrix effect interference by internal standard method.The accuracy of the method were good.The spiked recoveries of the detected elements were from 94.3% to 101.5%.The precision of the method were good (RSD < 3.5%,n =6).The linearities of the detected elements were good,and the correlation coefficients were all greater than 0.999.The detection limits were from 0.22 to 8.54 ng/mL.The quantization limits were from 0.68 to 25.73 ng/mL.Conclusion The established method was accurate,sensitivity,rapid and reliable,which can be applied to the determination of contents of 8 heavy metal elements in iron dextran.
ABSTRACT
Objective To establish a method for the determination of 8 heavy metal elements of Pb,Cd,As,Hg,Co,V,Se,Mo in iron dextran by inductively coupled plasma optical emission spectrometry (ICP-OES).Method Through selection of detection wavelengths,optimization of instrument parameters,verification of matrix effect interference and investigation of internal standard for correction of matrix effect interference,the 8 heavy metal elements were analyzed by ICP-OES.Results The recovery rate of the detected elements were increased from 83%-90% to 94%-100%,after correction of matrix effect interference by internal standard method.The accuracy of the method were good.The spiked recoveries of the detected elements were from 94.3% to 101.5%.The precision of the method were good (RSD < 3.5%,n =6).The linearities of the detected elements were good,and the correlation coefficients were all greater than 0.999.The detection limits were from 0.22 to 8.54 ng/mL.The quantization limits were from 0.68 to 25.73 ng/mL.Conclusion The established method was accurate,sensitivity,rapid and reliable,which can be applied to the determination of contents of 8 heavy metal elements in iron dextran.
ABSTRACT
Objective To establish a method for the determination of eight heavy metal elements ofPb,Cd,As,Hg,Co,V,Se,Mo in iron dextran by inductively coupled plasma optical emission spectrometry (ICP-OES).Method Through selection of detection wavelengths,optimization of instrument parameters,correction of spectral interference,verification of ionization interference and investigation of suppression methods,the eight heavy metal elements were analyzed by ICP-OES.Results The recovery rate of the detected elements increased 5% to 10% by suppression of ionization interference.The accuracy of the method was good.The spiked recovery rates of the detected elements were from 95.7% to 101.1%.The precision of the method were good (RSD < 3.6%,n =6).The linearities of the detected elements were good,and the correlation coefficients were all greater than 0.999.The detection limits were from 0.15 to 8.09 ng/mL.The quantization limits were from 0.46 to 24.26 ng/mL.Conclusion The method was accurate,sensitivity,rapid and reliable,which can be applied to the determination of contents of eight heavy metal elements in iron dextran.