1.
Acta bioquím. clín. latinoam ; 55(3): 303-309, jul. 2021. graf
Article in Spanish | LILACS, BINACIS | ID: biblio-1374053

ABSTRACT

Resumen El objetivo de este trabajo fue comparar los niveles de fibrinógeno (FBG) obtenidos por el método de Clauss con los obtenidos por el método de fibrinógeno derivado del tiempo de protrombina (FBG PT-d), con dos tromboplastinas, en pacientes anticoagulados con distintas drogas. Se estudiaron pacientes anticoagulados consecutivos: 105 con antagonistas de la vitamina K (AVK), 55 con heparina no fraccionada (HNF), 58 con heparina de bajo peso molecular (HBPM), 60 con rivaroxabán, 45 con apixabán, 60 con dabigatrán y 100 controles normales (CN). El FBG se determinó por el método de Clauss y FBG PT-d utilizando tromboplastina de cerebro de conejo o recombinante humana; los niveles de heparina, rivaroxabán y apixabán por método cromogénico anti Xa; el dabigatrán con el ensayo de tiempo de trombina diluido. Existió un sesgo positivo (p<0,001) al comparar el FBG PT-d vs. FBG por Clauss: CN: 13,7%, AVK: 31,8%, rivaroxabán: 34,8% y apixabán: 20,0% cuando se utilizó tromboplastina de conejo. En el caso de las muestras que contenían HBPM se observó este desvío con ambas tromboplastinas. El sesgo porcentual en presencia de dabigatrán y heparina no fraccionada no fue estadísticamente distinto del obtenido en el grupo control. El ensayo de FBG PT-d no debe utilizarse en pacientes anticoagulados con rivaroxabán, apixabán, HBPM o AVK, ya que sobreestima los niveles de FBG. El porcentaje de sesgo depende del tipo de tromboplastina utilizado y fue mayor con la de cerebro de conejo en el sistema de detección utilizado.


Abstract The aim of this study was to compare fibrinogen (FBG) results obtained by Clauss method (FBG-C) and by the prothrombin time-derived fibrinogen assay (FBG PT-d) with two thromboplastins in patients under anticoagulation. Consecutive anticoagulated patients were studied: 105 vitamin-K antagonist (VKA), 55 unfractioned heparin, 58 LMWH, 60 rivaroxaban, 45 apixaban and 60 dabigatran, and 100 healthy controls (NC). FBG-C was performed by Clauss and FIB PT-d with rabbit brain and human recombinant thromboplastins, respectively. Heparins, rivaroxaban and apixaban levels were measured by antiXa; dabigatran by thrombin diluted assay. A positive bias of FBG PT-d vs. FBG-C with both thromboplastins were seen in NC (13.7 and 19.0 % for HS and RP, respectively), but bias with HS in rivaroxaban, apixaban and VKA patients were significantly higher compared to NC: 34.8%, 20.0 % and 31.8 %, respectively. LMWH presented higher BIAS compared to NC with both thromboplastins. Samples with unfraction heparin and dabigatran presented similar bias to NC. FBG PT-d should not be used in patients under anticoagulant treatment because of an important overestimation of FBG could be obtained in these patients. The percentage of bias depends on the type of thromboplastin used; it was higher with rabbit brain thromboplastin in the detection system used.


Resumo O objetivo deste trabalho foi comparar os níveis de fibrinogênio (FBG) obtidos pelo método de Clauss com aqueles obtidos pelo método do fibrinogênio derivado do tempo de protrombina (FBG PT-d), com duas tromboplastinas, em pacientes anticoagulados com diferentes drogas. Pacientes anticoagulados consecutivos foram estudados: 105 com antagonista da vitamina K (AVK); 55 com heparina não fracionada (UFH); 58 com heparina de baixo peso molecular (HBPM), 60 com rivaroxabana, 45 com apixabana, 60 com dabigatrana e 100 controles normais (CN). FBG foi determinado pelo método de Clauss e FBG PT-d usando tromboplastina de cérebro de coelho ou tromboplastina humana recombinante; níveis de heparina, rivaroxabana e apixabana pelo método cromogênico anti-Xa; dabigatrana com ensaio de tempo de trombina diluída. Há um viés positivo (p<0,001) ao comparar o FBG PT-d vs FBG de Clauss: CN: 13,7%; AVK: 31,8%, rivaroxabana: 34,8% e apixabana 20,0% quando foi utilizada tromboplastina de coelho. No caso das amostras contendo HBPM, esse desvio foi observado com ambas as tromboplastinas. O viés percentual na presença de dabigatrana e heparina não fracionada não foi estatisticamente diferente daquela obtida no grupo controle. O ensaio de FBG PT-d não deve ser usado em pacientes anticoagulados com rivaroxabana, apixabana, LMWH ou VKA, pois superestima os níveis de FBG. A porcentagem de viés depende do tipo de tromboplastina utilizado e foi maior com a de cérebro de coelho, no sistema de detecção utilizado.


Subject(s)
Humans , Male , Female , Adult , Middle Aged , Aged , Fibrinogen/analysis , Prothrombin/administration & dosage , Blood Coagulation , Thromboplastin , Pharmaceutical Preparations/administration & dosage
2.
Rev. bras. anal. clin ; 51(1): 52-57, 30/03/2019. graf
Article in Portuguese | LILACS | ID: biblio-1008205

ABSTRACT

Objetivo: Analisar os testes de coagulação: tempo de protrombina (TP) e tempo de tromboplastina parcial (TTP) em diferentes tempos de centrifugação da amostra da biológica, com relação ao protocolo padrão do Clinical Laboratory Standards Institute (CLSI). Métodos: As amostras foram divididas em cinco alíquotas de 1 mL. Foi realizada a centrifugação em 15, 10, 5, 2 e 1 minuto, com velocidade de 1500 g. O TP e TTP foram imediatamente analisados em aparelho automatizado. Os plasmas foram analisados para presença de elementos residuais: eritrócitos, leucócitos e plaquetas. Resultados: Observou-se alteração dos valores do TP nos tempos de centrifugação 10, 5, 2 e 1 minuto e do TTP em 5, 2 e 1 minuto, com relação ao protocolo padrão. Na interpretação de Bland Altman, observou-se um viés significativo do limite clínico aceitável para o TP e para o TTP em todos os tempos de centrifugação, com relação ao protocolo padrão. Apenas no tempo de centrifugação de 15 minutos não foram encontradas células residuais nas amostras analisadas. Conclusão: O tempo de centrifugação de 15 minutos é o ideal para remoção completa das células sanguíneas residuais e para garantia da confiabilidade dos resultados dos testes de coagulação TP e TTP.


Objective: To analyze the coagulation tests: prothrombin test (PT) and partial thromboplastin time (PTT) in different centrifugation times of the sample, in relation to the standard protocol of the Clinical Laboratory Standards Institute (CLSI). Methods: The selected samples were splitted up into five aliquots of 1 mL. Centrifugation of these aliquots was carried out at 15, 10, 5, 2 and 1 minute at 1500 g. The PT and PTT were analyzed in an automated apparatus. The plasmas were analyzed for presence of residual elements: erythrocytes, leukocytes and platelets. Results: The results showed a change in the values of PT at the 10, 5, 2 and 1 minute centrifugation times and the PTT at 5, 2 and 1 minutes, relative to the standard protocol. In the interpretation of Bland Altman, a significant bias of the acceptable clinical limit for TP and TTP at all centrifugation times was observed, relative to the standard protocol. Only in the 15 minute centrifugation time no residual cells were found in the analyzed samples. Conclusion: The present study demonstrated that the 15-minute centrifugation time is ideal for complete removal of residual blood cells and to ensure the reliability of the results of the PT and PTT coagulation


Subject(s)
Humans , Male , Female , Prothrombin Time , Blood Coagulation Tests , Centrifugation
3.
Acta sci., Biol. sci ; 4020180000. ilus, tab, graf
Article in English | LILACS, VETINDEX | ID: biblio-1460788

ABSTRACT

The biotechnological value of macroalgae for screening assays of thrombin generation-TG using sulfated polysaccharides-SPs as substitutes to heparin has been poorly explored. Five Brazilian species of macroalgae (Gracilaria birdiae, Acanthophora muscoides, Halymenia sp., Caulerpa cupressoides and C. racemosa) wereanalyzed and compared for their abundance, physical-chemical characteristics and in vitro anticoagulant assays of activated partial thromboplastin time-APTT, prothrombin time-PT and TG. Papain extraction yielded (p 100 kDa. These procedures,combined with the use of Stains-All, also indicated nonSPs. APTTs ranged from 2.81 (A. muscoides) to 21.30 IU(Halymenia sp.) vs. heparin (193 IU), and were dependent on sulfation of the crude SPs. PT was not altered. Withrespect to TG assay, crude SPs modified concentration-dependent and independently from molecular mass TGby both intrinsic/extrinsic pathways in 60-fold diluted human plasma, with total intrinsic inactivation using crudeSPs from A. muscoides in parallel to heparin (p < 0.05). Thrombosis in vitro is differentially modulated by distinctcrude SPs from Brazilian seaweeds.


O valor biotecnológico das macroalgas para ensaios de varredura de geração de trombina-GT pouco tem sido explorado usando polissacarídeos sulfatados-PSs como substitutos à heparina. Foramanalisadas e comparadas cinco espécies brasileiras de macroalgas (Gracilaria birdiae, Acanthophora muscoides, Halymenia sp., Caulerpa cupressoides e C. racemosa) quanto à abundância, às característicasfísico-químicas e os ensaios anticoagulantes in vitro de tempo de tromboplastina parcial ativada-TTPA, aotempo de protrombina-TP e a GT. A extração com papaína rendeu (p 100 kDa. Esses procedimentos,combinados ao uso de azul de toluidina/Stains-All, indicaram também polissacarídeos-não sulfatados. OsTTPAs foram dependentes da sulfatação dos PSs brutos e variaram de 2,81 (A. muscoides) a 21,30 UI (Halymenia sp.) vs. heparina (193 UI). O TP não foi alterado. Com respeito ao ensaio de GT, os PSs brutos modificaram, dependente de concentração e independentemente de massa molecular, GT pelas viasintrínseca/extrínseca no plasma humano diluído 60 vezes, com inativação intrínseca total usando PSs brutosde A. muscoides em paralelo à heparina (p < 0,05). A trombose in vitro é modulada diferencialmente porPSs brutos distintos de algas marinhas brasileiras.


Subject(s)
Seaweed/enzymology , Seaweed/chemistry , Thrombin/analysis
4.
Acta bioquím. clín. latinoam ; 50(2): 181-192, jun. 2016. tab
Article in Spanish | LILACS | ID: biblio-837597

ABSTRACT

Los nuevos anticoagulantes orales directos (DOACs) son agentes sintéticos dirigidos en forma específica contra el factor Xa o contra la trombina. Por su farmacocinética predecible y escasa variabilidad interindividual no es necesario el monitoreo de anticoagulación. Esto es un cambio en el paradigma del tratamiento anticoagulante. Sin embargo, los DOACs alteran las pruebas de hemostasia y es importante saber cómo interpretarlas en caso de hemorragia, episodio trombótico o ante una cirugía de urgencia. Las pruebas clásicas de coagulación solo sirven para orientar sobre el efecto anticoagulante de los DOACs. Se debe conocer, para cada agente, cuál de las pruebas es útil, el reactivo que se utilizó y el tiempo transcurrido desde la ingesta del medicamento. Las pruebas específicas miden la concentración del DOAC. Son más sensibles, pero no han sido investigadas en relación a eventos clínicos y poseen una gran variación interindividual. Se concluye que hasta ahora no se cuenta con pruebas validadas para definir si se debe ajustar la dosis de los DOACs, aún en pacientes con alto riesgo hemorrágico. Nuevos estudios clínicos que utilizan diferentes pruebas de hemostasia serán de utilidad para ayudar a interpretar el efecto de los DOACs. Hasta que esto suceda, se tendrá que convivir con este nuevo paradigma, la ausencia del monitoreo.


New oral direct anticoagulants (DOACs) are synthetic agents targeting either thrombin or factor Xa. Their predictable pharmacokinetics and scarce variability eliminate the need for regular coagulation monitoring. This is a paradigm shift in anticoagulation treatment. However, DOACs interfere with coagulation assays and it is very important to recognize them in a bleeding or thrombotic event or in an urgent surgery. Commonly used screening coagulation tests are of limited utility and different test reagents, what kind of test to perform for each anticoagulant and the time of drug intake must be considered. Specific assays can accurately quantify drug levels, but they are not yet related to clinical events and there is great inter-individual variability. As a conclusion,there are no laboratory assays developed to support dose adjustment based on these test results. New clinical trials with laboratory measurement will help solve this. Until then, this new paradigm of no laboratory testing in routine clinical practice will have to be accepted.


Os novos anticoagulantes orais diretos (DOACs) são agentes sintéticos dirigidos em forma específica contra o fator Xa ou contra a trombina. Devido a sua farmacocinética previsível e escassa variabilidade interindividual não é necessária a monitoração de anticoagulação. Essa é uma mudança no paradigma do tratamento anticoagulante. Entretanto, os DOACs alteram as provas de hemostasia e é importante saber de que maneira interpretá-las em caso de hemorragia, episódio trombótico ou diante de uma cirurgia de urgência. Os testes clássicos de coagulação apenas servem para orientar a respeito do efeito anticoagulante dos DOACs. É necessário conhecer, para cada agente, qual dos testes é útil, o reagente que foi utilizado e o tempo decorrido desde a ingestão do medicamento. Os testes específicos medem a concentração do DOAC. São mais sensíveis, mas não foram pesquisados em relação a eventos clínicos e possuem grande variação interindividual. A conclusão é que até o momento não se conta com provas validadas para definir se deve ser ajustada a dosagem dos DOACs, mesmo em pacientes com alto risco hemorrágico. Novos estudos clínicos utilizando diferentes testes de hemostasia serão de utilidade para ajudar-nos a interpretar o efeito dos DOACs. Até que isto aconteça, será preciso conviver com este novo paradigma, a ausência da monitorização.


Subject(s)
Humans , Male , Female , Coagulants (Water Treatment) , Coagulation Agents , Anticoagulants , Thrombin , Hemostasis
5.
Acta bioquím. clín. latinoam ; 50(2): 291-301, jun. 2016. tab
Article in Spanish | LILACS | ID: biblio-837607

ABSTRACT

Los inhibidores adquiridos son defectos raros. Se asocian a diferentes manifestaciones clínicas con morbimortalidad significativa. Su detección es importante para implementar el tratamiento sin demora. Hay inhibidores específicos (bloquean función), contra todos los factores de la coagulación y además VWF, o de interferencia (afectan una o varias vías de la coagulación). Los inhibidores específicos son alo-anticuerpos desarrollados en pacientes deficitarios (complicación terapéutica) o auto-anticuerpos presentes en individuos sin alteraciones previas. Hay anticuerpos específicos que pueden afectar la depuración pero no inhiben función. Inhibidores de interferencia: inmunoglobulinas u otras sustancias (heparina/heparinoides, PDF/pdf, PIVKAS, moléculas anómalas, etc.) asociadas a diferentes situaciones clínicas (asintomáticos, sangrados, trombosis y/o complicaciones obstétricas). El laboratorio es fundamental para el diagnóstico. Las pruebas globales detectan el defecto, que no corrigen por el agregado de plasma normal; se caracteriza luego el tipo de inhibidor y eventualmente se titula. Esto es complejo; hay variabilidad en los resultados y posibilidad de falsos positivos o negativos, además las pruebas no son estrictamente específicas. Los algoritmos diagnósticos son útiles, pero no contemplan la posibilidad de defectos combinados. Es crítico caracterizar al inhibidor y descartar posibles interferencias o defectos concomitantes; ello requiere aplicar las pruebas adecuadas e interpretarlas correctamente.


Acquired inhibitors are rare disorders. They are associated with different clinical behaviours and significant morbi-mortality. Detection is important in order to start treatment urgently. There are either specific inhibitors, which block function, against all coagulation factors, and VWF, or with interference effects, on one or more coagulation pathways. Specific inhibitors are either allo-antibodies developed in deficient patients, which give rise to therapeutic complication; or auto-antibodies, which are present in individuals without previous defects. There are specific antibodies that can affect clearance but which cannot block the function. Inhibitors with interference effects are immunoglobulins or other substances (heparin/heparinoids, FDP/fdp, PIVKAS, abnormal molecules, etc.) associated with different clinical settings (asymptomatic, bleeding, thrombosis and/or obstetric complications). Laboratory results are fundamental for the diagnosis. Global tests are able to detect the defect, which is not corrected by the addition of normal plasma; the type of inhibitor is then characterized and titration of the inhibitor is eventually performed. This is complex; there is variability in the results and there is likelihood of false positive or negative results; moreover, the tests are not strictly specific. Diagnostic algorithms are useful tools but they do not consider combined defects. It is a critical point to characterize the inhibitor and exclude possible interferences or concomitant defects; this demands application of the correct tests without misinterpretations.


Os inibidores adquiridos são defeitos raros. Associam-se a diferentes manifestações clínicas com morbimortalidade significativa. Sua detecção é importante para implementar o tratamento sem demora. Há inibidores específicos (bloqueiam função), contra todos os fatores da coagulação e também VWF, ou de interferência (afetam uma ou várias vias da coagulação). Inibidores específicos: - aloanticorpos desenvolvidos em pacientes deficitários (complicação terapêutica); - autoanticorpos, em indivíduos sem alterações prévias. Existem anticorpos específicos que podem afetar a depuração, mas não inibem função. Inibidores de interferência: imunoglobulinas ou outras substâncias (heparina/heparinoides, PDF/pdf, PIVKAS, moléculas anômalas, etc.) associadas a diferentes situações clínicas (assintomáticos, sangramentos, tromboses e/ou complicações obstétricas). O laboratório é fundamental para o diagnóstico. Os testes globais detectam o defeito, que não corrige pelo acréscimo de plasma normal; caracteriza-se depois o tipo de inibidor e eventualmente é titulado. Isto é complexo; há variabilidade nos resultados e possibilidade de falsos positivos ou negativos, além disso os testes não são rigorosamente específicos. Os algoritmos diagnósticos são úteis, mas não consideram a possibilidade de defeitos combinados. É crítico caracterizar o inibidor e descartar possíveis interferências ou defeitos concomitantes; isso requer aplicar os testes adequados e interpretá-los corretamente.


Subject(s)
Humans , Male , Female , Blood Coagulation , Lupus Coagulation Inhibitor , Antibodies , Clinical Laboratory Techniques , Enzyme Inhibitors
6.
São Paulo; s.n; 2015. 118 p. ilus, tab, graf.
Thesis in Portuguese | LILACS | ID: biblio-972063

ABSTRACT

INTRODUÇÃO: sangramento por coagulopatia é problema comum durante o transplante hepático (TH). O uso adequado da monitorização da coagulação pode reduzir a transfusão de hemocomponentes, como o Plasma Fresco Congelado (PFC). Exames Convencionais da Coagulação (ECC), tais como Tempo de Protrombina (TP) e Tempo de Tromboplastina Parcial Ativada (TTPa), são os testes mais amplamente utilizados para monitorizar a coagulação durante o TH, mas algumas limitações têm sido apontadas acerca do seu uso em pacientes cirróticos. OBJETIVO: investigar o uso de ECC como variáveis preditoras da indicação de transfusão de PFC durante o TH em pacientes cirróticos. MÉTODO: analisou-se coorte histórica de 297 transplantes hepáticos com enxertos provenientes de doadores cadáveres. Foram incluídos receptores cirróticos de uma única instituição durante nove anos (2002-2010). A infusão profilática de ácido épsilon-aminocaproico (20 mg/kg/h) e outros pré-requisitos hemostáticos foram mantidos na cirurgia. O TP [expresso na forma de Percentual de Atividade da Protrombina (TP%) e de Relação Normalizada Internacional (INR)] e o TTPa foram medidos no pré-operatório e no fim de cada fase do TH. Os participantes só receberam transfusão de PFC quando se diagnosticou coagulopatia, independentemente dos resultados dos ECC. Os pacientes foram distribuídos em dois grupos, de acordo com a ocorrência de transfusão intraoperatória de PFC. Examinou-se o comportamento dos resultados dos ECC durante a cirurgia. Analisaram-se os fatores de risco para a transfusão de PFC por análises uni e multivariada. Os resultados pós-operatórios de ambos os grupos foram comparados. A acurácia dos ECC para predizer o uso de PFC em cada fase da cirurgia foi investigada por curvas ROC. Além disso, pontos de corte dos ECC não associados à coagulopatia foram calculados para cada fase da cirurgia...


BACKGROUND & AIMS: Bleeding due to coagulopathy is a common problem during liver transplantation (LT). Coagulation monitoring may reduce transfusion of blood components, including Fresh Frozen Plasma (FFP). Conventional coagulation assays (CCA), like Prothrombin Time (PT) and Activated Partial Thromboplastin Time (aPTT), are the most widely employed tests to monitor coagulation during LT, but some limitations have been assigned to their use in cirrhotic patients. This study investigated the predictive value of these blood coagulation tests in predicting FFP transfusions during LT in cirrhotic patients. METHODS: This historical cohort study analyzed 297 isolated, deceased donor LTs performed in cirrhotic patients from a single institution during a nine-year period (2002 - 2010). Prophylactic infusion of epsilon-aminocaproic acid [EACA] (20 mg/kg/h) and other hemostatic requirements were maintained intraoperatively. PT [expressed as Activity Percentage (PT%) and International normalized ratio (INR)] and aPTT [expressed in seconds] were measured preoperatively and by the end of each phase of LT. Hemostatic blood components were transfused only in case of coagulopathy. Patients were divided in two groups according to intraoperative FFP transfusion: FFP group and Non-FFP group. Behavior of CCA results during LT were examined in both groups. Univariate and multivariate analyses of risk factors associated with FFP transfusion were performed. Post-operative outcomes were compared between groups. Accuracy of CCA to predict FFP transfusions was investigated using receiver operating characteristic (ROC) curves. Also, alert values of CCA unassociated with coagulopathy in each phase of surgery were calculated...


Subject(s)
Humans , Liver Cirrhosis , Liver Transplantation , Blood Coagulation Disorders , Blood Coagulation Tests , Predictive Value of Tests , Blood Component Transfusion
7.
Arq. neuropsiquiatr ; 72(7): 487-489, 07/2014.
Article in English | LILACS | ID: lil-714593

ABSTRACT

The use of oral anticoagulation treatment (OAT) in patients with an international normalized ratio (INR) higher than 1.7 is a contraindication to thrombolysis in acute ischemic stroke. The aim of the present study is to compare the use of point-of-care (POC) coagulometers to the standard coagulation analysis (SCA) procedure of the INR as a decision-making test for use with patients taking OAT. Method: Eighty patients on chronic OAT underwent a POC and an SCA during a regular outpatient evaluation. Results: When comparing the abilities of the POC test and the SCA test to identify adequate levels for thrombolysis (≤1.7), the POC had a sensitivity of 96.6% (95%CI 88.4-99.1) and a specificity of 60.0% (95%CI 38.6-78). POC overestimated INR levels by 0.51 points compared to the SCA test. Conclusion: POC has a high sensitivity compared to the SCA test for the identification of patients within the cut-off point for thrombolysis. .


Valores acima de 1.7 no RNI de pacientes em uso de terapia anticoagulante (ACO) é uma contraindicação para a trombólise no AVC isquêmico. O Objetivo do presente estudo é comparar o teste capilar rápido (TCR) com o exame tradicional de RNI como método de decisão para indicação de terapia trombolítica em pacientes em uso de ACO. Método: Oitenta pacientes em uso crônico de ACO foram submetidos ao TCR e ao exame tradicional durante consulta ambulatorial. Resultados: Ao serem comparados os resultados do TCR com o exame tradicional com o ponto de corte de 1,7 de RNI, o teste rápido teve sensibilidade de 96,6% (IC95% 88,4-99,1) e especificidade de 60,0% (IC95% 38,6-78). O TCR superestimou os níveis de RNI em 0,51 pontos comparado ao exame tradicional. Conclusão: O TCR apresentou elevada sensibilidade, comparado ao exame tradicional para a identificação de pacientes em uso de ACO com indicação de trombólise. .


Subject(s)
Aged , Female , Humans , Male , Middle Aged , Anticoagulants/therapeutic use , International Normalized Ratio/standards , Point-of-Care Systems/standards , Thrombolytic Therapy , Brain Ischemia/drug therapy , International Normalized Ratio/methods , Reference Values , Reproducibility of Results , Sensitivity and Specificity , Statistics, Nonparametric , Stroke/drug therapy
8.
UNOPAR Cient., Ciênc. biol. saude ; 16(3): 229-237, jul. 2014. ilus
Article in Portuguese | LILACS-Express | LILACS | ID: lil-720050

ABSTRACT

Exames laboratoriais são importantes na definição do diagnóstico, quando este não pode ser esclarecido exclusivamente pelos dados de história clínica e exame físico. Compete ao Cirurgião-Dentista solicitar exames complementares, pois alguns problemas sistêmicos poderão complicar e alterar o prognóstico ou, até mesmo, contraindicar a realização de cirurgias ou procedimentos invasivos. Este tema é relevante, pois grande porcentagem de profissionais formados em Odontologia apresenta dificuldade na solicitação e interpretação de exames laboratoriais. Este estudo tem como objetivo revisar conceitos básicos de interpretação e de critérios para solicitação de exames laboratoriais como: Hemograma e Coagulograma completo. O presente trabalho foi desenvolvido por meio da análise documental da produção bibliográfica obtida por meio da base de dados: Biblioteca Virtual em Saúde (BVS), Biblioteca Digital Brasileira de Teses e Dissertações (BDTD) e na base Periódicos CAPES. Concluímos que os exames laboratoriais, quando bem indicados, colaboram para decidirmos quais as condutas que devemos tomar. Pois, diante dos valores encontrados, estando acima ou abaixo dos valores de referência, o profissional os interpretando de forma correta, pode prevenir situações de infecções secundárias, má cicatrização, hemorragias e complicações no tratamento odontológico. Sendo assim, profissionais que solicitam e sabem interpretar exames laboratoriais estão oferecendo maior segurança ao seu paciente.


Laboratory tests are important in defining the diagnosis, when this cannot be explained by data from clinical history and physical examination. Dentists must request exams because some systemic problems may complicate and alter the prognosis or even contraindicate performing surgeries or invasive procedures. This topic is relevant because a large percentage of graduates in dentistry have difficulty ordering and interpretation of laboratory tests. This study aims to review basic concepts of interpretation and criteria for requesting laboratory tests such as complete blood count and Coagulation. This study was developed through documentary analysis of bibliographic production obtained through the database: Virtual Health Library (VHL), Brazilian Digital Library of Theses and Dissertations (BDTD) and CAPES Journals. We conclude that the laboratory tests, when properly indicated, collaborate to decide the correct procedure. For the values found before, being above or below the reference values, when interpreting them correctly, the professional can prevent situations of secondary infections, poor wound healing, bleeding and complications during dental treatment. Thus, professional that interpret laboratory tests correctly can offer greater safety to the patient.

9.
Rev. bras. anal. clin ; 43(3): 237-244, 2011. tab, graf
Article in Portuguese | LILACS | ID: lil-651511

ABSTRACT

O Sysmex® CA-1500 é um analisador automatizado de coagulação que detecta a formação de fibrina por princípio ótico.A avaliação do seu desempenho analítico foi realizada através de testes de precisão, reprodutibilidade, arraste e linearidade, além de ensaios para verificar a influência de fatores pré-analíticos como relação anticoagulante / sangue e hemólise nas determinações deTP, TTP, TT e fibrinogênio. Além disso, os resultados de TP e TTP foram correlacionados com os obtidos no coagulômetro ST4, Diagnostica STAGO, que utiliza o método eletromagnético para a detecção do coágulo de fibrina. Observou-se grande variação dosresultados sobre a influência da hemólise (TP: 0,91 a 2,33%; TTP: 1,37 a 6,54% e TT: 1,38 a 12,72%) e relação anticoagulante/sangue (TP: 1,39 a 659,3%; TTP: 2,93 a 275,2%, TT: 4,0 a 128% e fibrinogênio: 3,4 a 32,2%), quando comparados com amostras sem hemólise e com a relação anticoagulante /sangue de 1:9. A porcentagem de arraste foi (TP: 0,75%, TTP: 0,50% e fibrinogênio: 5,65%). Aprecisão intraensaio para TP, TTP, TT e fibrinogênio obteve CV < 3% para plasmas normais e patológicos, assim como a reprodutibilidade interensaios em plasma normal liofilizado e pool de plasmas de pacientes normais. As correlações para TP e TTP entre os doiscoagulômetros avaliados obteve valores de r ≥ 0,90. A linearidade foi boa até aproximadamente 650 mg/dL de fibrinogênio e obteve r = 0,981. Os resultados obtidos demonstraram que o CA-1500 realiza os testes de coagulação com desempenho analítico satisfatório.


The Sysmex® CA-1500 is an automated coagulation analyzer that detects the formation of fibrin clot by optical principle.The evaluation of its analytical performance was accomplished through tests of precision, reproducibility, linearity and carryover, as well as tests to check the influence of pre-analytical factors such as the anticoagulant / blood ratio and hemolysis on the determinations of PT, PTT, TT and fibrinogen. Moreover, the results of PT and PTT were correlated with those obtained in the ST4 coagulometer, Diagnostica Stago, using the eletromagnetic method for the detection of fibrin clot. We found great variation inresults under the influence of hemolysis (PT: 0,91 to 2,33%, PTT: 1,37 to 6,54% and TT: 1,38 to 12,72%) and anticoagulant / blood ratio (PT: 1,39 to 659,3%, PTT: 2,93 to 275,2%, TT: 4,0 to 128% and fibrinogen: 3,4 to 32,2%) compared with samples withouthemolysis and the ratio of 1:9 anticoagulant / blood. The percentage of carryover was (PT: 0,75%, PTT: 0,50% and fibrinogen: 5,65%). The intra-assay precision for PT, PTT, TT and Fibrinogen had CV < 3% for normal and abnormal plasmas, as well as interassayreproducibility in normal lyophilized plasma and normal pooled plasma. The correlations between TP and TTP on the two evaluated coagulometers obtained r values ≥ 0,90. The linearity was good up to about 650 mg/dl of fibrinogen and obtained r = 0,981. The results showed that CA-1500 perform coagulation tests with satisfactory analytical performance.


Subject(s)
Humans , Adult , Automation, Laboratory , Automation, Laboratory/instrumentation , Quality Control , Blood Coagulation Tests/methods