1.
Yeungnam University Journal of Medicine ; : 143-150, 1998.
Article in Korean | WPRIM | ID: wpr-96003

ABSTRACT

The aim of this study was to evaluate domestic enzyme immunoassay(EIA) kit ?LG HCD 3.0?(LG) for the detection of antibody to hepatitis C virus(anti-HCV) in comparision with Axsym HCV version 3.0(Axsym), Cobas Core anti-HCV EIA(Cobas). Cobas kit shows better clear distinction between positive and negative by signal/cutoff ratio(S/C), but it also reveal relatively high false positive rate. The concordance rate of test results between LG and Axsym was 96.2%, between LG and Cobas was 95.5%, and total agreement between three EIA kit was 93.9%. LG were relative poor distinction between positive and negative results, but it could be applied clinically as a screening tool for hepatitis C in general population. The S/C of one false negative result by LG was 0.91, and false positive were less than 4.0, therefore we concluded it is necessary to confirm by immunoblotting assay when S/C were between 0.8 and 4.0.


Subject(s)
Hepacivirus , Hepatitis C , Immunoblotting , Immunoenzyme Techniques , Mass Screening
SELECTION OF CITATIONS
SEARCH DETAIL