1.
Journal of Experimental Hematology ; (6): 851-855, 2022.
Article in Chinese | WPRIM | ID: wpr-939699

ABSTRACT

OBJECTIVE@#Compared with the method of optical microscopy, to evaluate the accuracy of fragmented red cells(FRC) detection by Sysmex XN-3000.@*METHODS@#A total of 111 samples were collected from patients diagnosed as thrombotic thrombocytopenic purpura, autoimmune disease, hematological disease, malignant tumor and health examination in our hospital from June 2019 to February 2021, including 74 cases in the case group and 37 cases in the healthy control group. All samples were detected by optical microscope and Sysmex XN-3000, respectively. ROC was used to evaluate the detection ability of Sysmex XN-3000 for schistocyte. Bland-Altman method was used to evaluate the consistency of the results of the two methods for detection of schistocyte, and Pearson correlation analysis was conducted for the difference of the results.@*RESULTS@#The area under the ROC curve was 0.890(95% CI: 0.828-0.952, P<0.01). Sysmex XN-3000 count did not quantitatively agree with schistocyte counts by microscopy in the case group(mean of difference:-1.53, 95% limits of agreement: -8.78~5.72). There was a weak positive correlation between platelet count and the difference of analyzer and microscopic results (r=0.32,P<0.05).@*CONCLUSION@#Sysmex XN-3000 can be used as a reference for qualitative determination of schistocyte. However, the sensitivity of Sysmex XN-3000 should be improved. It is still necessary to combine with manual microscopy. The quantitative results are not reliable now and cannot be used as a reference for monitoring the results of schistocyte in clinical patients after treatment.


Subject(s)
Humans , Neoplasms , Platelet Count , Purpura, Thrombotic Thrombocytopenic , ROC Curve , Reproducibility of Results
2.
Chinese Journal of Clinical Laboratory Science ; (12): 166-170, 2018.
Article in Chinese | WPRIM | ID: wpr-694820

ABSTRACT

Objective To evaluate the creditability of warning message of white differential count (WDF) and white precursor cell (WPC) channels in Sysmex XN-3000 hematology analyzer,and verify its optimal threshold and adjust the alarm threshold.Methods A total of 61 EDTA-K2 anticoagulated blood samples without abnormal warning and 521 EDTA-K2 anticoagulated blood samples with abnormal warning were simultaneously detected in WDF and WPC channels.After the smear specimens of blood sample were automatically prepared by the instrument,microscopic examinations were performed manually.The results of microscopic examination were considered as the gold standard to determine the reliability of the warning message from the instrument and verify the reasonability of initial warning threshold value provided by the manufacture.Consequently,the threshold values were adjusted based on the requirements in practical work.Results The warning messages of atypical lymphocytes and blasts/abnormal lymphocytes in WDF channel were higher sensitive (95.8% and 100% respectively),but lower specific (34.7% and 23.5% respectively) compared with microscopic examination.The warning messages of atypical lymphocyte,blasts and abnormal lymphocytes in WPC channel were lower sensitive (81.3%,66.7%,and 76.5% respectively) but higher specific (61.9%,55.5% and 88.3 % respectively) compared with microscopic examination.According to the ROC curve analysis,the prognostic values of warning message of microscopic examination were of medium level,except the warning message for abnormal lymphocytes was poor compared with WPC channel.Combining the practical retest rules,the optimal critical threshold values of atypical lymphocytes and blasts/Abn lymph in WDF channel were adjusted as 120,and they were adjusted as 140 in WPC channel.Conclusion The high sensitive WDF channel should first be used for screening,and the detectable warning message could be retested by using high specific WPC channel to shorten the turnaround time of the test results and improve the working efficiency.The initial critical warning threshold provided by the manufacture should be verified and adjusted to the optimum critical threshold in order to ensure the accuracy of test results.

3.
Journal of Laboratory Medicine and Quality Assurance ; : 42-46, 2017.
Article in Korean | WPRIM | ID: wpr-156755

ABSTRACT

Platelet distribution width (PDW) is an index for platelet size variation. In this study, we analysed the correlation between PDW values obtained using two different hematology analysers that employ different measurement methods. Complete blood cell parameters including PDW for 153 healthy individuals were measured using both, ADVIA 2120i (Simens AG, Germany) and XN-3000 (Sysmex, Japan). The PDW values measured using the two hematology analysers showed a moderate correlation (r=0.661, P0.900, P<0.001). PDW obtained using XN-3000 showed a strong correlation with mean platelet volume, whereas PDW obtained using ADVIA 2120i did not. The reference values in this group were 40.0%–64.2% in ADVIA 2120i and 9.0–16.0 fL in XN-3000. In conclusion, PDW values obtained using ADVIA 2120i and XN-3000 are not interchangeable. In laboratories equipped with more than one hematology analyser, a particular analyser should be used consistently for monitoring a particular patient.


Subject(s)
Humans , Blood Cells , Blood Platelets , Erythrocyte Indices , Erythrocytes , Hematology , Leukocytes , Mean Platelet Volume , Platelet Count , Reference Values
4.
China Medical Equipment ; (12): 38-40,41, 2016.
Article in Chinese | WPRIM | ID: wpr-604533

ABSTRACT

Objective:To evaluate the performance of reticulocyte (Ret) detected by Sysmex XN-3000 automatic blood analyzer.Methods: 96 blood samples were collected randomly in clinical detection of reticulocytes, and respectively were detected by XN-3000 automatic blood analyzer and microscopy. The precision, stability, linearity and carryover of Ret in blood samples detected by XN-3000 automatic blood analyzer were evaluated by the correlation analysis of the results from the two methods and the correlation analysis would be completed by microscopy.Results: precision, linear range and sensitivity of Ret measured by XN-3000 were within the allowable range. The stability did not change significantly below 4℃ within 48h (CV<1%). The carryover rate was 0.14%. The results measured by XN-3000 were well compared with microscopy (r=0.979).Conclusion: Sysmex XN-3000 automatic blood analyzer is applied to the clinical detection of large quantities of specimens and it is ideal for measuring Ret.