1.
Korean Journal of Medical Physics ; : 33-41, 2012.
Artículo en Inglés | WPRIM | ID: wpr-81825

RESUMEN

For applying the quality assurance (QA) of volumetric modulated arc therapy (VMAT) introduced in Eulji Hospital, we classify it into three different QA steps, treatment planning QA, pretreatment delivering QA, and treatment verifying QA. These steps are based on the existing intensity modulated radiation therapy (IMRT) QA that is currently used in our hospital. In each QA step, the evaluated items that are from QA program are configured and documented. In this study, QA program is not only applied to actual patient treatment, but also evaluated to establish a reference of clinical acceptance in pretreatment delivering QA. As a result, the confidence limits (CLs) in the measurements for the high-dose and low-dose regions are similar to the conventional IMRT level, and the clinical acceptance references in our hospital are determined to be 3 to 5% for the high-dose and the low-dose regions, respectively. Due to the characteristics of VMAT, evaluation of the intensity map was carried out using an ArcCheck device that was able to measure the intensity map in all directions, 360degrees. With a couple of dosimetric devices, the gamma index was evaluated and analyzed. The results were similar to the result of individual intensity maps in IMRT. Mapcheck, which is a 2-dimensional (2D) array device, was used to display the isodose distributions and gave very excellent local CL results. Thus, in our hospital, the acceptance references used in practical clinical application for the intensity maps of 360degrees directions and the coronal isodose distributions were determined to be 93% and 95%, respectively. To reduce arbitrary uncertainties and system errors, we had to evaluate the local CLs by using a phantom and to cooperate with multiple organizations to participate in this evaluation. In addition, we had to evaluate the local CLs by dividing them into different sections about the patient treatment points in practical clinics.


Asunto(s)
Humanos , Radioterapia de Intensidad Modulada
2.
Korean Journal of Medical Physics ; : 206-215, 2011.
Artículo en Coreano | WPRIM | ID: wpr-153492

RESUMEN

The purpose of this was to investigate the measurement of fluence dose map for the specific patient quality assurance. The measurement of fluence map was performed using 2D matrixx detector. The absorbed dose was measured by a glass detector, Gafchromic film and ion chamber in Hybrid Optimized VMAT Phantom (HOVP). For 2D Matrixx, the results of comparison were average passing rate 85.22%+/-1.7 (RT_Target), 89.96%+/-2.15 (LT_Target) and 95.14%+/-1.18 (G4). The dose difference was 11.72%+/-0.531, -11.47%+/-0.991, 7.81%+/-0.857, -4.14%+/-0.761 at the G1, G2, G3, G4. In HOVP, the results of comparison for film were average passing rate (3%, 3 mm) 93.64%+/-3.87, 90.82%+/-0.99. We were measured an absolute dose in steep gradient area G1, G2, G3, G4 using the glass detector. The difference between the measurement and calculation are 8.3% (G1), -5.4% (G2), 6.1% (G3), 7.2% (G4). The using an Ion-chamber were an average relative dose error -1.02%+/-0.222 (Rt_target), 0.96%+/-0.294 (Lt_target). Though we need a more study using a transmission detector. However, a measurement of real-time fluence map will be predicting a dose for real-time specific patient quality assurance in volume modulated arc therapy.


Asunto(s)
Humanos , Quimera , Vidrio
3.
Korean Journal of Medical Physics ; : 216-223, 2011.
Artículo en Coreano | WPRIM | ID: wpr-153491

RESUMEN

Recently PTW developed a MicroLion liquid ionization chamber which is water_equivalent and has a small sensitive volume of 0.002 cm3. The aim of this work is to investigate such dosimetric characteristics as dose linearity, dose rate dependency, spatial resolution, and output factors of the chamber for the external radiotherapy photon beam. The results were compared to those of Semiflex chamber, Pinpoint chamber and Diode chamber with the sensitive volumes of 0.125 cm3, 0.03 cm3 and 0.0025 cm3, respectively and evaluated to be suitable for small fields. This study was performed in the 6MV photon energy from a Varian 2300 C/D linac accelerator and the MP3 water phantom (PTW, Freiburg) was used. Penumbras in the varios field sizes ranged from 0.5x0.5 cm2 to 10x10 cm2 were used to evaluate the spatial resolution. Output factors were measured in the field sizes of 0.5x0.5 to 40x40 cm2. Readings of the chamber was linearly proportional to dose. Dose rate dependency was measured from 100 MU/min to 600 MU/min, showed a maximum difference of 5.0%, and outputs decreased with dose rates. The spatial resolutions determined with comparing profiles for the field sizes of 0.5x0.5 cm2 to 10x10 cm2 agreed between every detector except the Semiflex chamber to within 2%. Outputs of detectors were compared to that of Semiflex chamber and showed good agreements within 2% for every chamber. This study shows that MicroLion chamber characterized by a high signal-to-noise ratio and water equivalence could be suitable for the small field dosimetry.


Asunto(s)
Dependencia Psicológica , Lectura , Relación Señal-Ruido , Análisis Espacial , Agua
4.
Korean Journal of Medical Physics ; : 120-125, 2010.
Artículo en Coreano | WPRIM | ID: wpr-30096

RESUMEN

For the measurements of an absorbed dose using the standard dosimetry based on an absorbed dose to water the variety of factors, whether big, small, or tiny, may influence the accuracy of dosimetry. The beam quality correction factor kappa(Q, Q(0))of an ionization chamber might also be one of them. The cylindrical type of ionization chamber, the PTW30013 chamber, was chosen for this work and 9 chambers of the same type were collected from several institutes where the chamber types are used for the reference dosimetry. They were calibrated from the domestic Secondary Standard Dosimetry Laboratory with the same electrometer and cable. These calibrated chambers were used to measure absorbed doses to water in the reference condition for the photon beam of 6 MV and 10 MV and the electron beam of 12 MeV from Siemens ONCOR. The biggest difference among chambers amounts to 2.4% for the 6 MV photon beam, 0.8% for the 10 MV photon beam, and 2.4% for the 12 MeV electron beam. The big deviation in the photon of 6 MV demonstrates that if there had been no problems with the process of measurements application of the same kappa(Q, Q(0)) to the chambers used in this study might have influenced the deviation in the photon 6 MV and that how important an external audit is.


Asunto(s)
Academias e Institutos , Electrones , Agua
5.
Korean Journal of Medical Physics ; : 223-231, 2010.
Artículo en Coreano | WPRIM | ID: wpr-55608

RESUMEN

We have taken surveys about total 72 departments of radiation oncology which is performing the treatment with linear accelerator and brachytherapy unit in Korea. The survey was included the research about the linear accelerator, brachytherapy, Also, we surveyed the various performance (QA period, manpower, time) of quality control for understanding of efficiency. The survey results show that the QA test of daily and weekly are almost same comparing to USA and Europe but the QA performance of monthly and yearly in Korea are 15.5 which is less than USA and Europe recommended QA item number of 17 to 21. The manpower and QA time in Korea also lower than 50% of USA and Europe recommended because the manpower and QA time limitation in Korea. It will be expected that the manual of quality management in each clinic could be appropriately established when combining the present results with previously published AAPM TG-40 and other protocols.


Asunto(s)
Braquiterapia , Europa (Continente) , Corea (Geográfico) , Aceleradores de Partículas , Control de Calidad , Oncología por Radiación
6.
Korean Journal of Medical Physics ; : 97-105, 2009.
Artículo en Coreano | WPRIM | ID: wpr-115852

RESUMEN

Absorbed dose to water based protocols recommended that plane-parallel chambers be calibrated against calibrated cylindrical chambers in a high energy electron beam with R50>7 g/cm2 (E> or =16 MeV). However, such high-energy electron beams are not available at all radiotherapy centers. In this study, we are compared the absorbed dose to water determined according to cross-calibration method in a high energy electron beam of 16 MeV and in electron beam energies of 12 MeV below the cross-calibration quality remark. Absorbed dose were performed for PTW 30013, Wellhofer FC65G Farmer type cylindrical chamber and for PTW 34001, Wellhofer PPC40 Roos type plane-parallel chamber. The cylindrical and the plane-parallel chamber to be calibrated are compared by alternately positioning each at reference depth, zref=0.6R50-0.1 in water phantom. The DW of plane-parallel chamber are derived using across-calibration method at high-energy electron beams of 16, 20 MeV. Then a good agreement is obtained the DW of plane-parallel chamber in 12 MeV. The agreement between 20 MeV and 12 MeV are within 0.2% for IAEA TRS-398.


Asunto(s)
Electrones , Agua
7.
Korean Journal of Medical Physics ; : 317-323, 2009.
Artículo en Coreano | WPRIM | ID: wpr-227380

RESUMEN

The standard dosimetry systems based on an absorbed dose to water recommend to use a planeparallel chamber for the calibration of such a low-megavoltage electron beam as a nominal energy of 6 MeV. For this energy ranges of an electron beam a cylindrical chamber should not be used for the routinely regular beam calibration, but the feasibility of the temporary use of a cylindrical chamber was studied to give temporary solutions for special situations users meet. The PTW30013 chambers and the electron beam quality of R(50)=2.25 g/cm2 were selected for this study. 10 PTW30013 chambers, a cylindrical type of chamber, were calibrated in KFDA, the secondary standards dosimetry laboratories, and given the absorbed dose-to-water calibration factors, respectively. A "temporary" kappa(Q,Q0) for each chamber were calculated using the absorbed dose determined by a cross-calibrated planeparallel chamber, with the result of an average 0.9352 for 10 chambers. This value for PTW30013 chamber was used to determine an absorbed dose to water at the reference depth. The absorbed doses determined by PTW30013 chambers were in an agreement within 2% with that by ROOS chamber. In a certain situation where a cylindrical chamber be used instead of a planeparellel chamber, the value of 0.9352 might be useful to determine an absorbed dose to water in the same beam quality of electron beam as this study.


Asunto(s)
Calibración , Electrones , Agua
8.
Korean Journal of Medical Physics ; : 139-144, 2009.
Artículo en Coreano | WPRIM | ID: wpr-137645

RESUMEN

Recently, the mathematical analysis of stent simulation has been improved, with the help of development of various tool which measure mechanical property and location of stent in artery. The most crucial part of the stent modeling is how to design ideal stent and to evaluate the interaction between stent and artery. While there has been great deal of researches on the evaluation of the expansion, stress distribution, deformation of the stent in terms of the various parameters, few verification through computer simulation has been performed about deformation and stress distribution of the stent. In this study, we have produced the corresponding results between experimental test using Universal Testing Machine and computer simulation for the ideal model of stent. Also, we have analyzed and compared stress distribution of stent in the cases of that with membrane and that without membrane. The results of this study would provide minimum change of plan and good quality for ideal stent replacing damaged artery through the analysis using computer simulation in the early stage of stent design.


Asunto(s)
Aleaciones , Arterias , Colodión , Simulación por Computador , Prueba de Esfuerzo , Análisis de Elementos Finitos , Membranas , Stents
9.
Korean Journal of Medical Physics ; : 139-144, 2009.
Artículo en Coreano | WPRIM | ID: wpr-137644

RESUMEN

Recently, the mathematical analysis of stent simulation has been improved, with the help of development of various tool which measure mechanical property and location of stent in artery. The most crucial part of the stent modeling is how to design ideal stent and to evaluate the interaction between stent and artery. While there has been great deal of researches on the evaluation of the expansion, stress distribution, deformation of the stent in terms of the various parameters, few verification through computer simulation has been performed about deformation and stress distribution of the stent. In this study, we have produced the corresponding results between experimental test using Universal Testing Machine and computer simulation for the ideal model of stent. Also, we have analyzed and compared stress distribution of stent in the cases of that with membrane and that without membrane. The results of this study would provide minimum change of plan and good quality for ideal stent replacing damaged artery through the analysis using computer simulation in the early stage of stent design.


Asunto(s)
Aleaciones , Arterias , Colodión , Simulación por Computador , Prueba de Esfuerzo , Análisis de Elementos Finitos , Membranas , Stents
10.
Korean Journal of Medical Physics ; : 167-173, 2009.
Artículo en Coreano | WPRIM | ID: wpr-137637

RESUMEN

As the radiotherapy technique development, the needs for using of medical electronic chart in the department of radiation oncology is growing. However, the complexity of affairs of radiation oncology make it difficult to develop a electronic medical chart. In this study, we introduce the electronic medical chart developed by domestic hospital. The function and example of electronic medical chart designed as radiation treatment progress was showed and the future study was presented.


Asunto(s)
Electrónica , Electrones , Oncología por Radiación
11.
Korean Journal of Medical Physics ; : 167-173, 2009.
Artículo en Coreano | WPRIM | ID: wpr-137636

RESUMEN

As the radiotherapy technique development, the needs for using of medical electronic chart in the department of radiation oncology is growing. However, the complexity of affairs of radiation oncology make it difficult to develop a electronic medical chart. In this study, we introduce the electronic medical chart developed by domestic hospital. The function and example of electronic medical chart designed as radiation treatment progress was showed and the future study was presented.


Asunto(s)
Electrónica , Electrones , Oncología por Radiación
12.
Korean Journal of Medical Physics ; : 174-179, 2009.
Artículo en Coreano | WPRIM | ID: wpr-137635

RESUMEN

To track moving tumor in real time, CyberKnife system imports a technique of the synchrony respiratory tracking system. The fiducial marker which are detectable in X-ray images were demand in CyberKnife Robotic radiosurgery system. It issued as reference markers to locate and track tumor location during patient alignment and treatment delivery. Fiducial marker implantation is an invasive surgical operation that carries a relatively high risk of pneumothorax. Most recently, it was developed a direct lung tumor registration method that does not require the use of fiducials. The purpose of this study is to measure the accuracy of target applying X-sight lung tracking using the Gafchromic film in dynamic moving thorax phantom. The X-sight Lung Tracking quality assurance motion phantom simulates simple respiratory motion of a lung tumor and provides Gafchromic dosimetry film-based test capability at locations inside the phantom corresponding to a typical lung tumor. The total average error for the X-sight Lung Tracking System with a moving target was 0.85+/-0.22 mm. The results were considered reliable and applicable for lung tumor treatment in CyberKnife radiosurgery system. Clinically, breathing patterns of patients may vary during radiation therapy. Therefore, additional studies with a set real patient data are necessary to evaluate the target accuracy for the X-sight Lung Tracking system.


Asunto(s)
Humanos , Marcadores Fiduciales , Pulmón , Neumotórax , Radiocirugia , Respiración , Tórax , Atletismo
13.
Korean Journal of Medical Physics ; : 174-179, 2009.
Artículo en Coreano | WPRIM | ID: wpr-137634

RESUMEN

To track moving tumor in real time, CyberKnife system imports a technique of the synchrony respiratory tracking system. The fiducial marker which are detectable in X-ray images were demand in CyberKnife Robotic radiosurgery system. It issued as reference markers to locate and track tumor location during patient alignment and treatment delivery. Fiducial marker implantation is an invasive surgical operation that carries a relatively high risk of pneumothorax. Most recently, it was developed a direct lung tumor registration method that does not require the use of fiducials. The purpose of this study is to measure the accuracy of target applying X-sight lung tracking using the Gafchromic film in dynamic moving thorax phantom. The X-sight Lung Tracking quality assurance motion phantom simulates simple respiratory motion of a lung tumor and provides Gafchromic dosimetry film-based test capability at locations inside the phantom corresponding to a typical lung tumor. The total average error for the X-sight Lung Tracking System with a moving target was 0.85+/-0.22 mm. The results were considered reliable and applicable for lung tumor treatment in CyberKnife radiosurgery system. Clinically, breathing patterns of patients may vary during radiation therapy. Therefore, additional studies with a set real patient data are necessary to evaluate the target accuracy for the X-sight Lung Tracking system.


Asunto(s)
Humanos , Marcadores Fiduciales , Pulmón , Neumotórax , Radiocirugia , Respiración , Tórax , Atletismo
14.
Korean Journal of Medical Physics ; : 7-13, 2009.
Artículo en Inglés | WPRIM | ID: wpr-88373

RESUMEN

This work is for the preliminary study for the calibration of an (192)Ir brachytherapy source based on an absorbed dose to water standards. In order to calibrate brachytherapy sources based on absorbed dose to water standards using a clyndirical ionization chamber, the beam quality correction factor kappa(Q,Q0) is needed. In this study kappa(Q,Q0)s were determined by both Monte carlo simulation and semiexperimental methods because of the realistic difficulties to use primary standards to measure an absolute dose at a specified distance. The 5 different serial numbers of the PTW30013 chamber type were selected for this study. While chamber to chamber variations ran up to maximum 4.0% with the generic kappa gen(Q,Q0), the chamber to chamber variations were within a maximum deviation of 0.5% with the individual kappa ind(Q,Q0). The results show why and how important ionization chambers must be calibrated individually for the calibration of (192)Ir brachytherapy sources based on absorbed dose to water standards. We hope that in the near future users will be able to calibrate the brachytherapy sources in terms of an absorbed dose to water, the quantity of interest in the treatment, instead of an air kerma strength just as the calibration in the high energy photon and electron beam.


Asunto(s)
Braquiterapia , Calibración , Electrones , Agua
15.
Korean Journal of Medical Physics ; : 14-20, 2009.
Artículo en Inglés | WPRIM | ID: wpr-88372

RESUMEN

We measured the dose distribution for spinal cord and tumor using Gafchromic film, applying 3D and 4D-Treatment Planning for lung tumor within the phantom. A measured dose distribution was compared with a calculated dose distribution generated from 3D radiation treatment planning and 4D radiation treatment planning system. The agreement of the dose distribution in tumor for 3D and 4D treatment planning was 90.6%, 97.64% using gamma index computed for a distance to agreement of 1 mm and a dose difference of 3%. However, a gamma agreement index of 3% dose difference tolerence of and 2 mm distance to agreement, the accordance of the dose distribution around cord for 3D and 4D radiation treatment planning was 57.13%, 90.4%. There are significant differences between a calculated dose and a measured dose for 3D radiation treatment planning, no significant differences for 4D treatment planning. The results provide the effectiveness of the 4D treatment planning as compared to 3D. We suggest that the 4-dimensional treatment planning should be considered in the case where such equipments as Cyberknife with the real time tracking system are used to treat the tumors in the moving organ.


Asunto(s)
Pulmón , Radiocirugia , Médula Espinal , Atletismo
16.
The Journal of the Korean Society for Therapeutic Radiology and Oncology ; : 218-227, 2009.
Artículo en Inglés | WPRIM | ID: wpr-21051

RESUMEN

PURPOSE: To investigate the radiobiologic effects of neutron and X-ray irradiation on DU-145 prostate carcinoma cells by identifying the differences of HIF-1alpha expression and apoptosis. MATERIALS AND METHODS: Nude mice were injected with the human prostate cancer cell line, DU-145, and then irradiated with 2 Gy and 10 Gy X-rays, or 0.6 Gy and 3.3 Gy neutrons, respectively. The mice were sacrificed at 24 hours and 120 hours after irradiation. The expression levels of HIF-1alpha, Bcl-2 and Bax were compared with immunohistochemical staining and western blotting. The apoptotic indexes were compared with the Terminal deoxynucleotidyl biotin-dUTP nick and labeling (TUNEL) assay. RESULTS: At day 1, HIF-1alpha and Bcl-2 expression decreased, while Bax expression and the number of TUNEL positive cells increased in neutron irradiated groups for the control and X-ray irradiated groups. The Bcl-2/Bax ratio was significantly lower in the neutron irradiated groups regardless of dose (p=0.001). The same pattern of the differences in the expressions of the HIF-1alpha, Bcl-2, Bax, Bcl-2/Bax ratio, and apoptotic indexes were indentified at day 5. HIF-1alpha expression was related with Bcl-2 (p=0.031), Bax (p=0.037) expressions and the apoptotic indexes (p=0.016) at day 5. CONCLUSION: Neutron irradiation showed a decrease in HIF-1alpha, Bcl-2 expression, and Bcl-2/Bax ratio, but increased Bax expression regardless of dose. This study suggests that the differences radiobiological responses between photon and neutron irradiation may be related to different HIF-1alpha expression and subsequent apoptotic protein expressions.


Asunto(s)
Animales , Humanos , Ratones , Apoptosis , Western Blotting , Línea Celular , Etiquetado Corte-Fin in Situ , Ratones Desnudos , Neutrones , Próstata , Neoplasias de la Próstata , Trasplantes
17.
Korean Journal of Medical Physics ; : 305-312, 2008.
Artículo en Coreano | WPRIM | ID: wpr-93126

RESUMEN

The aim of this study is to introduce the accuracy of Ir-192 source's apparent activity using the well-type chamber and the Farmer-type ionization chamber in the high dose rate brachytherapy. We measured the apparent activity of Ir-192 that each medical center in the country has and the apparent activity of calibration certificate provided by manufacturer is compared with that by our experimental measurement. The number of sources used for the activity comparison was 5. The accuracy of the measured activity was in the range of -2.8% to -1.0% and -2.1% to 0.2% for the Farmer-type chamber system (Jig) and for the well-type, respectively. The maximum difference was within 1.0% for comparison with two calibration's tool. Our results demonstrate that well-type chamber as wall as Farmer-type chamber is a appropriate system as the routine source calibration procedures in HDR brachytherapy. Whenever a new source is installed to use in clinics, by periods, a source calibration should be carried out.


Asunto(s)
Braquiterapia , Calibración
18.
Korean Journal of Medical Physics ; : 13-19, 2007.
Artículo en Coreano | WPRIM | ID: wpr-27797

RESUMEN

The purpose of study is to investigate whether glass rod detector (GRD) would be suitable for dosimeter of radiotherapy units. A GRD is used for the measurement of the output factors and x-axis beam profiles from Gamma Knife. The output factors measured with GRD from the 14, 8 and 4 mm collimators relative to the 18 mm collimator are 0.980+/-0.013, 0.949+/-0.013 and 0.872+/-0.012, respectively. The output factors obtained with a GRD are within 1.0% in good agreement with the values recommended by the manufacture. The full width at half maximum (FWHM) of x-axis beam profiles measured with GRD are 5.9 mm at a 4 mm collimator.


Asunto(s)
Vidrio , Radioterapia
19.
The Journal of the Korean Society for Therapeutic Radiology and Oncology ; : 138-143, 2006.
Artículo en Coreano | WPRIM | ID: wpr-104901

RESUMEN

PURPOSE: To analyze the long-term stability of Farmer-type cylindrical ionization chambers by calibration factor provided from the KFDA (Korea Food Drug Administration) MATERIALS AND METHODS: The cylindrical ionization chambers used in this study were the PTW 30001 (30006), 30013, 30002, 30004, 23333, the Capintec PR06C, the NE 2571, the Exradin A12 and the Wellhofer FC65G (IC70). We were analyzed that the N(k) and N(D,W) calibration factor for the cylindrical chambers and compared between the measured N(D,W) and calculated N(D,W) calibration factor. RESULTS: We have observed that the long-term stability of the PTW 30013 (30006), the Wellhofer FC65G (IC70) and the NE 2571 has varied within 0.2%. The measured N(D,W) calibration factor was about 1.0% higher than the calculated N(D,W) that determined by the N(k) calibration factor. CONCLUSION: The study has evaluated that the long-term stability of the cylindrical chambers through analysis for the Nk and N(D,W) calibration factor. It has contributed to the improvement of clinical electron dosimetry in radiotherapy centers.


Asunto(s)
Calibración , Radioterapia
20.
The Journal of the Korean Society for Therapeutic Radiology and Oncology ; : 185-191, 2006.
Artículo en Coreano | WPRIM | ID: wpr-53723

RESUMEN

PURPOSE: Standardization quality assurance (QA) program of CyberKnife for suitable circumstances in Korea has not been established. In this research, we investigated the development of QA program for CyberKnife and evaluation of the feasibility under applications. MATERIALS AND METHODS: Considering the feature of constitution for systems and the therapeutic methodology of CyberKnife, the list of quality control (QC) was established and divided dependent on the each period of operations. And then all these developed QC lists were categorized into three groups such as basic QC, delivery specific QC, and patient specific QC based on the each purpose of QA. In order to verify the validity of the established QA program, this QC lists was applied to two CyberKnife centers. The acceptable tolerance was based on the undertaking inspection list from the CyberKnife manufacturer and the QC results during last three years of two CyberKnife centers in Korea. The acquired measurement results were evaluated for the analysis of the current QA status and the verification of the propriety for the developed QA program. RESULTS: The current QA status of two CyberKnife centers was evaluated from the accuracy of all measurements in relation with application of the established QA program. Each measurement result was verified having a good agreement within the acceptable tolerance limit of the developed QA program. CONCLUSION: It is considered that the developed QA program in this research could be established the standardization of QC methods for CyberKnife and confirmed the accuracy and stability for the image-guided stereotactic radiotherapy.


Asunto(s)
Humanos , Constitución y Estatutos , Corea (Geográfico) , Prácticas Mortuorias , Control de Calidad , Radioterapia