Annals of Laboratory Medicine ; : 121-124, 2013.
Artículo en Inglés | WPRIM | ID: wpr-216012

RESUMEN

Three automated immunoassay kits for anti-Hepatitis A Virus (HAV) IgM-Architect, (Abbott Laboratories, USA), Elecsys (Roche Diagnostics, Germany), and ADVIA Centaur (Siemens Healthcare Diagnostics Inc., USA)-were compared. We included 178 consecutive samples, for which an anti-HAV IgM test was requested at Seoul National University Hospital from September 2009 to January 2010. Reviewing of medical records, reverse transcription (RT)-PCR for HAV RNA, or total anti-HAV assay were performed on 16 (9.0%) samples with discrepant results. The percent agreements (kappas) of the Architect and ADVIA Centaur, Architect and Elecsys, and ADVIA Centaur and Elecsys kits were 96.6% (0.91), 96.6% (0.92), and 97.8% (0.94), respectively. Eight out of 16 discrepant samples showed gray-zone values in Architect but were nonreactive in the others. Slightly earlier seroconversion was suspected in Elecsys. The 3 assays showed comparable performances with excellent agreements in a tertiary care hospital setting.


Asunto(s)
Humanos , Virus de la Hepatitis A/genética , Hepatitis B/diagnóstico , Hospitales Universitarios , Inmunoensayo/métodos , Inmunoglobulina M/análisis , ARN Viral/análisis , Juego de Reactivos para Diagnóstico , Estudios Retrospectivos , Reacción en Cadena de la Polimerasa de Transcriptasa Inversa , Atención Terciaria de Salud