Journal of the Korean Society of Magnetic Resonance in Medicine ; : 116-122, 2001.
Artículo en Inglés | WPRIM | ID: wpr-10129

RESUMEN

Purpose : To find sensitivity of MRI imaging methods to slow flow phantom study was performed with conventional Spin-Echo, gradient echo based Phase Contrast, fast GLASS, and heavily T2-weighted Fast Spin Echo pulse sequences. MATERIALS AND METHODS: A siphon driven flow phantom was constructed with a ventriculo-peritoneal shunt catheter and a GE phantom to achieve continuous variable flow. Four different pulse sequences including Spin-Echo, Phase Contrast, GRASS and Heavily T2-weighted Fast Spin Echo were evaluated to depict slow flow in the range from 0.08 ml/min to 1.7 ml/min and to compare signal intensities between static fluid and flowing fluid. RESULTS: In the slow flow above 0.17 ml/min conventional Spin-Echo showed superior apparent contrast between static and flowing fluid while GRASS was more sensitive to the very slow flow below 0.17 ml/min. It was not accurate to calculate flow and velocity below 0.1 ml/min with a modified PC imaging. CONCLUSION: Four different MR pulse sequences demonstrated different sensitivity to the range of slow flow from 0.08 ml/min to 1.7 ml/min. This finding may be clinically useful to measure CSF shunt flow or detecting CSF collection and thrombosis.


Asunto(s)
Catéteres , Vidrio , Imagen por Resonancia Magnética , Poaceae , Trombosis , Derivación Ventriculoperitoneal