Arq. bras. cardiol ; 76(6): 437-444, June 2001. ilus, tab
Article Dans Portugais, Anglais | LILACS | ID: lil-286361

Résumé

OBJECTIVE: This study was performed to observe the number of pacemakers that had never been reprogrammed after implantation, and the effect of optimised output programming on estimated longevity of pulse generators in patients with pacemaker METHODS: Sixty patients with Teletronics Reflex pacemakers were evaluated in a pacemaker clinic, from the time of the beginning of its activities, in June 1998, until March 1999. Telemetry was performed during the first clinic visit, and we observed how many pulse generators retained nominal output settings of the manufactures indicating the absence of reprogramming until that date. After evaluation of the capture threshold, reprogramming of pacemakers was performed with a safety margin of 2 to 2.5:1, and we compared the estimated longevity based on battery current at the manufacturer's settings with that based on settings achieved after reprogramming. RESULTS: In 95 percent of the cases, the original programmed setting was never reprogrammed before the patients attended the pacemaker clinic. Reprogramming the pacemaker prolonged estimated pulse generator life by 19.7Ý15.6 months (35.5 percent). CONCLUSION: The majority of the pacemakers evaluated had never been reprogrammed. Estimated pulse generator longevity can be prolonged significantly, using this simple, safe, efficacious, and cost-effective procedure


Sujets)
Mâle , Adulte d'âge moyen , Humains , Femelle , Alimentations électriques , Pacemaker , Logiciel , Pacemaker/économie , Télémétrie , Facteurs temps
SÉLECTION CITATIONS
Détails de la recherche