1.
International Eye Science ; (12): 921-924, 2007.
Artigo em Chinês | WPRIM | ID: wpr-641669

RESUMO

· AIM: To compare the measurement of anterior chamber depth (ACD) and axial length (AL) by IOLMaster and contact ultrasonic (US) axial scan (A-scan).· METHODS: Measurements of ACD and AL were prospectively obtained in 137 eyes of 121 subjects with the IOLMaster compared with measurements with the US.· RESULTS: There was an excellent correlation between IOLMaster and US measurements for the ACD (r=0.823;P<0.001) and AL (r=0.996;P<0.001). The mean values of the parameters measured by IOLMaster and US were,respectively, as follows: ACD, 2.94±0.49mm, 2.58±0.51mm;AL, 24.37±3.04mm, 23.81±2.83mm. The mean differences of ACD and AL values between IOLMaster and US measurements were 0.36 ±0.30mm, 0.56 ±0.34 mm respectively, and they proved to be statistically significant (P<0.001), With the 95%limits of agreement (LoA) from -0.08mm to +0.38mm for ACD and from -0.09mm to +0.69mm for AL.· CONCLUSION: As noncontact biometry, IOLMaster provides accurate values. A high degree of agreement between US and IOLMaster was noted. It not only has the advantage of performing noncontact examinations, but also produces various additional data simultaneously and may thus obviate the need for multiple examinations. Further studies are needed to assess the interchangeability of measurements in clinical practice.

2.
International Eye Science ; (12): 537-539, 2006.
Artigo em Chinês | WPRIM | ID: wpr-641745

RESUMO

AIM: To assess the relative agreement of GAT and NCT in IOP measurement by comparing the differences between Goldmann applanation tonometer (GAT) and non-contact tonometer (NCT) in intraocular pressure (IOP) detection.METHODS: IOP of 529 eyes of 265 volunteers were measured with both NCT and GAT, respectively.RESULTS: The measurement results of NCT were lower than that of GAT, there was significant difference between the IOP measured with NCT and GAT (19.13 vs23.43, t=22.644, P<0.05). With the increasing of IOP values, the difference magnitude was greater, especially in IOP group that was more than 30mmHg, but the correlation coefficient became lower.CONCLUSION: The measurement results with NCT are lower than that of GAT. When the IOP with the NCT is in borderline value, it need be corrected with GAT, in order to discover the pathologically elevated IOP and avoid the misdiagnosis and mistreatment of glaucoma.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA