Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Adicionar filtros








Intervalo de ano
1.
Chinese Journal of Medical Instrumentation ; (6): 201-230, 2007.
Artigo em Chinês | WPRIM | ID: wpr-323285

RESUMO

<p><b>OBJECTIVE</b>To establish a method of determining the residual epoxyethane in medical instruments.</p><p><b>METHOD</b>An HP-5 capillary column (30 m x 0.32 mm x 0.25 microm), an FID detecter and an oven were used in the determination. The oven temperature was kept at 120 degrees C for 8 minutes, the example inlet temperature was kept at 200 degrees C, while the detecter temperature was kept at 250 degrees C.</p><p><b>RESULT</b>The method has a linear range between 5.1920-51.920 microg, (r = 0.997, n = 6). The average rate of recovery of the injecter is 103.3%. The average rate of recovery of the infusion devices is 81.04%.</p><p><b>CONCLUSION</b>The method has been proved to be convenient, sensitive and accurate. It is suitable for the determination of residual organic solvents in medical instruments.</p>


Assuntos
Cromatografia Gasosa , Métodos , Equipamentos Descartáveis , Compostos de Epóxi , Etano , Solventes