1.
China Pharmacy ; (12): 3428-3430, 2016.
Artigo em Chinês | WPRIM | ID: wpr-504928

RESUMO

OBJECTIVE:To a evaluation method for the measurement uncertainty for the content of Bisacodyl enteric-coated tablet. METHODS:HPLC external standard method was conducted for content determination of Bisacodyl enteric-coated tablet, and mathematical model for uncertainty evaluation was established to systematically analyze and evaluate the influential factors in processes of solution preparation and instrument measurement. RESULTS:HPLC external standard method showed the content was 97.8%,confidence probability was 95%,expanded uncertainty was 2.8%,and determination result was (97.8 ± 2.8)%,k=2. CONCLUSIONS:The established method is suitable for the evaluation of measurement uncertainty for the content of Bisacodyl en-teric-coated tablet. Regularly calibrated verification for HPLC equipment and strict control of the weighing process will help to im-prove the accuracy measured by HPLC.