Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Adicionar filtros








Intervalo de ano
1.
EMHJ-Eastern Mediterranean Health Journal. 2008; 14 (6): 1365-1371
em Inglês | IMEMR | ID: emr-157280

RESUMO

This cross-sectional study aimed to investigate if the Tono-Pen, Schiotz and Perkins tonometers could be used interchangeably in general practice for measuring elevated intraocular pressure. A total of 74 eyes of 37 paediatric patients under general anaesthesia were checked with all 3 tonometers. All of the tonometers gave significantly different measurements from each other. However, with a mean difference of 1.4 mmHg and 95% limits of agreement of -5.7 to +8.6, the greatest agreement was between the Perkins and Tono-Pen tonometers. The Perkins tonometer is a hand-held variant of the Goldmann tonometer [the gold standard for intraocular pressure measures]. Therefore the Tono-Pen with its ease of use and safety could be a reliable device for use in general practice


Assuntos
Feminino , Humanos , Masculino , Pressão Intraocular/instrumentação , Pressão Intraocular/análise , Anestesia Geral , Glaucoma/diagnóstico , Estudos Transversais , Diagnóstico Precoce , Fatores de Risco
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA