Telehealth application of an ultrasonic home spirometer.
Doumit, Michael; Ledwos, Rianna; Plush, Leanne; Chuang, Sandra; Gray, Melinda; Jaffe, Adam; McBride, Jamie.
  • Doumit M; Department of Health Sciences, Macquarie University, Sydney, New South Wales, Australia michael.doumit@mq.edu.au.
  • Ledwos R; Physiotherapy, Sydney Children's Hospitals Network Randwick, Randwick, New South Wales, Australia.
  • Plush L; Respiratory Medicine, Sydney Children's Hospitals Network Randwick, Randwick, New South Wales, Australia.
  • Chuang S; Respiratory Medicine, Sydney Children's Hospitals Network Randwick, Randwick, New South Wales, Australia.
  • Gray M; Respiratory Medicine, Sydney Children's Hospitals Network Randwick, Randwick, New South Wales, Australia.
  • Jaffe A; School of Women's and Children's Health, University of New South Wales, Sydney, New South Wales, Australia.
  • McBride J; Respiratory Medicine, Sydney Children's Hospitals Network Randwick, Randwick, New South Wales, Australia.
Arch Dis Child ; 107(8): 752-754, 2022 08.
Artículo en Inglés | MEDLINE | ID: covidwho-1950043
ABSTRACT

OBJECTIVE:

To investigate the validity and home use of a personal ultrasonic spirometer.

METHODS:

Supervised spirometry was performed using laboratory equipment and a personal ultrasonic spirometer. In addition, the ability of children to perform acceptable spirometry during supervised telehealth appointments at home was assessed.

RESULTS:

59 children completed spirometry on both devices. There was high between-device intraclass correlation coefficient (ICC) for forced expiratory volume in 1 s (FEV1) and forced vital capacity (FVC) ICC 0.991 (95% CI 0.985 to 0.995) and 0.989 (95% CI 0.981 to 0.993), respectively. Bland-Altman analysis revealed mean bias and limits of agreement of -0.01 (-0.22 to 0.24) L for FEV1 and -0.02 (-0.30 to 0.33) L for FVC. 125 of 140 (89%) supervised telehealth spirometry sessions were acceptable.

CONCLUSION:

There was excellent reliability in between-device measurements; however, the limits of agreement were wide. Therefore, caution is needed if the device is used interchangeably with laboratory equipment. High success rates of telehealth spirometry sessions indicate the device is suitable for this application.