Your browser doesn't support javascript.
Daily SARS-CoV-2 Nasal Antigen Tests Miss Infected and Presumably Infectious People Due to Viral Load Differences among Specimen Types.
Viloria Winnett, Alexander; Akana, Reid; Shelby, Natasha; Davich, Hannah; Caldera, Saharai; Yamada, Taikun; Reyna, John Raymond B; Romano, Anna E; Carter, Alyssa M; Kim, Mi Kyung; Thomson, Matt; Tognazzini, Colten; Feaster, Matthew; Goh, Ying-Ying; Chew, Yap Ching; Ismagilov, Rustem F.
  • Viloria Winnett A; California Institute of Technology, Pasadena, California, USA.
  • Akana R; California Institute of Technology, Pasadena, California, USA.
  • Shelby N; California Institute of Technology, Pasadena, California, USA.
  • Davich H; California Institute of Technology, Pasadena, California, USA.
  • Caldera S; California Institute of Technology, Pasadena, California, USA.
  • Yamada T; Pangea Laboratory LLC, Tustin, California, USA.
  • Reyna JRB; Zymo Research Corporation, Irvine, California, USA.
  • Romano AE; Pangea Laboratory LLC, Tustin, California, USA.
  • Carter AM; California Institute of Technology, Pasadena, California, USA.
  • Kim MK; California Institute of Technology, Pasadena, California, USA.
  • Thomson M; California Institute of Technology, Pasadena, California, USA.
  • Tognazzini C; California Institute of Technology, Pasadena, California, USA.
  • Feaster M; Pasadena Public Health Department, Pasadena, California, USA.
  • Goh YY; Pasadena Public Health Department, Pasadena, California, USA.
  • Chew YC; Pasadena Public Health Department, Pasadena, California, USA.
  • Ismagilov RF; Pangea Laboratory LLC, Tustin, California, USA.
Microbiol Spectr ; : e0129523, 2023 Jun 14.
Artículo en Inglés | MEDLINE | ID: covidwho-20244296
ABSTRACT
In a recent household transmission study of SARS-CoV-2, we found extreme differences in SARS-CoV-2 viral loads among paired saliva, anterior nares swab (ANS), and oropharyngeal swab specimens collected from the same time point. We hypothesized these differences may hinder low-analytical-sensitivity assays (including antigen rapid diagnostic tests [Ag-RDTs]) by using a single specimen type (e.g., ANS) from reliably detecting infected and infectious individuals. We evaluated daily at-home ANS Ag-RDTs (Quidel QuickVue) in a cross-sectional analysis of 228 individuals and a longitudinal analysis (throughout infection) of 17 individuals enrolled early in the course of infection. Ag-RDT results were compared to reverse transcription-quantitative PCR (RT-qPCR) results and high, presumably infectious viral loads (in each, or any, specimen type). The ANS Ag-RDT correctly detected only 44% of time points from infected individuals on cross-sectional analysis, and this population had an inferred limit of detection of 7.6 × 106 copies/mL. From the longitudinal cohort, daily Ag-RDT clinical sensitivity was very low (<3%) during the early, preinfectious period of the infection. Further, the Ag-RDT detected ≤63% of presumably infectious time points. The poor observed clinical sensitivity of the Ag-RDT was similar to what was predicted based on quantitative ANS viral loads and the inferred limit of detection of the ANS Ag-RDT being evaluated, indicating high-quality self-sampling. Nasal Ag-RDTs, even when used daily, can miss individuals infected with the Omicron variant and even those presumably infectious. Evaluations of Ag-RDTs for detection of infected or infectious individuals should be compared with a composite (multispecimen) infection status to correctly assess performance. IMPORTANCE We reveal three findings from a longitudinal study of daily nasal antigen rapid diagnostic test (Ag-RDT) evaluated against SARS-CoV-2 viral load quantification in three specimen types (saliva, nasal swab, and throat swab) in participants enrolled at the incidence of infection. First, the evaluated Ag-RDT showed low (44%) clinical sensitivity for detecting infected persons at all infection stages. Second, the Ag-RDT poorly detected (≤63%) time points that participants had high and presumably infectious viral loads in at least one specimen type. This poor clinical sensitivity to detect infectious individuals is inconsistent with the commonly held view that daily Ag-RDTs have near-perfect detection of infectious individuals. Third, use of a combination nasal-throat specimen type was inferred by viral loads to significantly improve Ag-RDT performance to detect infectious individuals.
Palabras clave

Texto completo: Disponible Colección: Bases de datos internacionales Base de datos: MEDLINE Tipo de estudio: Estudio de cohorte / Estudios diagnósticos / Estudio experimental / Estudio observacional / Estudio pronóstico / Ensayo controlado aleatorizado Tópicos: Variantes Idioma: Inglés Revista: Microbiol Spectr Año: 2023 Tipo del documento: Artículo País de afiliación: Spectrum.01295-23

Similares

MEDLINE

...
LILACS

LIS


Texto completo: Disponible Colección: Bases de datos internacionales Base de datos: MEDLINE Tipo de estudio: Estudio de cohorte / Estudios diagnósticos / Estudio experimental / Estudio observacional / Estudio pronóstico / Ensayo controlado aleatorizado Tópicos: Variantes Idioma: Inglés Revista: Microbiol Spectr Año: 2023 Tipo del documento: Artículo País de afiliación: Spectrum.01295-23