Comparative evaluation of Luminex based assays for detection of SARS-CoV-2 antibodies in a transplantation laboratory.
Cox, A; Stevens, M; Kallon, D; Gupta, A; White, E.
  • Cox A; Clinical Transplantation Laboratory, 3rd Floor Pathology & Pharmacy Building, 80 Newark Street, London E1 2ES, United Kingdom; The University of Manchester, Oxford Road, Manchester M13 9PL, United Kingdom. Electronic address: amber.marie.c@sky.com.
  • Stevens M; Immunology Laboratory, Royal Sussex County Hospital Barry, Eastern Rd, Brighton BN2 5BE, United Kingdom.
  • Kallon D; Clinical Transplantation Laboratory, 3rd Floor Pathology & Pharmacy Building, 80 Newark Street, London E1 2ES, United Kingdom.
  • Gupta A; Clinical Transplantation Laboratory, 3rd Floor Pathology & Pharmacy Building, 80 Newark Street, London E1 2ES, United Kingdom.
  • White E; Clinical Transplantation Laboratory, 3rd Floor Pathology & Pharmacy Building, 80 Newark Street, London E1 2ES, United Kingdom.
J Immunol Methods ; 517: 113472, 2023 06.
Artículo en Inglés | MEDLINE | ID: covidwho-2293701
ABSTRACT

BACKGROUND:

Detection of SARS-CoV-2 antibodies is essential in establishing the parameters of an individual's immune response to COVID-19, from both natural infection and vaccination. Despite this, there is currently limited clinical guidance or recommendations for serological methods for their measurement. Here, we evaluate and compare four Luminex-based assays for the multiplex detection of IgG SARS-CoV-2 antibodies.

METHODS:

The four assays tested were Magnetic Luminex Assay, MULTICOV-AB Assay, Luminex xMAP SARS-CoV-2 Multi-Antigen IgG Assay and LABScreen COVID Plus Assay. Each assay's ability to detect antibodies to SARS-CoV-2 Spike (S), Nucleocapsid (N) and Spike-Receptor Binding Domain (RBD) was evaluated using 50 test samples (25 positive, 25 negative), previously tested by a widely used ELISA technique.

RESULTS:

The MULTICOV-AB Assay had the highest clinical performance detecting antibodies to S trimer and RBD in 100% (n = 25) of known positive samples. Both the Magnetic Luminex Assay and LABScreen COVID Plus Assay showed significant diagnostic accuracy with sensitivities of 90% and 88% respectively. The Luminex xMAP SARS-CoV-2 Multi-Antigen IgG Assay demonstrated limited detection of antibodies to the S antigen resulting in a sensitivity of 68%.

CONCLUSION:

Luminex-based assays provide a suitable serological method for multiplex detection of SARS-CoV-2 specific antibodies, with each assay able to detect antibodies to a minimum of 3 different SARS-CoV-2 antigens. Assay comparison identified there is moderate performance variability between manufacturers and further inter-assay variation of antibodies detected to different SARS-CoV-2 antigens.
Asunto(s)
Palabras clave

Texto completo: Disponible Colección: Bases de datos internacionales Base de datos: MEDLINE Asunto principal: COVID-19 Tipo de estudio: Estudios diagnósticos / Estudio experimental / Estudio pronóstico Tópicos: Vacunas Límite: Humanos Idioma: Inglés Revista: J Immunol Methods Año: 2023 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS


Texto completo: Disponible Colección: Bases de datos internacionales Base de datos: MEDLINE Asunto principal: COVID-19 Tipo de estudio: Estudios diagnósticos / Estudio experimental / Estudio pronóstico Tópicos: Vacunas Límite: Humanos Idioma: Inglés Revista: J Immunol Methods Año: 2023 Tipo del documento: Artículo