[Reuse process of positive pressure powered air-filter protective hoods].
Mo, Junjun; Huang, Fang; Lyu, Bei; Shen, Hongmei; Wang, Qun; Xu, Xiaowei; Chai, Qinming.
  • Mo J; Central Sterile Supply Department, the First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou 310003, China.
  • Huang F; Central Sterile Supply Department, the First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou 310003, China.
  • Lyu B; Central Sterile Supply Department, the First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou 310003, China.
  • Shen H; Central Sterile Supply Department, the First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou 310003, China.
  • Wang Q; Central Sterile Supply Department, the First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou 310003, China.
  • Xu X; Central Sterile Supply Department, the First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou 310003, China.
  • Chai Q; Central Sterile Supply Department, the First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou 310003, China.
Zhejiang Da Xue Xue Bao Yi Xue Ban ; 49(5): 603-608, 2020 Oct 25.
Artículo en Chino | MEDLINE | ID: covidwho-934531
ABSTRACT

OBJECTIVE:

To establish reuse process of positive pressure powered air-filter protective hoods during coronavirus disease 2019 (COVID-19) epidemic.

METHODS:

The procedure of pretreatment, storage, recovery, cleaning, disinfection and sterilization process of positive pressure powered air-filter protective hoods, which were used in the treatment of COVID-19 infection patients was established in Central Sterile Supply Department of the hospital. The cleaning and disinfection effects of the protective hoods after treatment were examined by magnifying glass method, residual protein detection method, real-time PCR, and agar pour plate method.

RESULTS:

Twenty five used protective hoods underwent totally 135 times of washing, disinfecting and sterilizing procedures. After washing, all the protein residue tests and COVID-19 nucleic acid tests showed negative results. After sterilizing, all the protective hoods met sterility requirement. All the tested protective hoods were undamaged after reprocessing.

CONCLUSIONS:

The established reuse procedures for used positive pressure powered air-filter protective hoods are safe.
Asunto(s)
Palabras clave

Texto completo: Disponible Colección: Bases de datos internacionales Base de datos: MEDLINE Asunto principal: Neumonía Viral / Esterilización / Desinfección / Equipo Reutilizado / Infecciones por Coronavirus / Pandemias / Filtros de Aire Idioma: Chino Revista: Zhejiang Da Xue Xue Bao Yi Xue Ban Asunto de la revista: Medicina Año: 2020 Tipo del documento: Artículo País de afiliación: J.issn.1008-9292.2020.10.09

Similares

MEDLINE

...
LILACS

LIS


Texto completo: Disponible Colección: Bases de datos internacionales Base de datos: MEDLINE Asunto principal: Neumonía Viral / Esterilización / Desinfección / Equipo Reutilizado / Infecciones por Coronavirus / Pandemias / Filtros de Aire Idioma: Chino Revista: Zhejiang Da Xue Xue Bao Yi Xue Ban Asunto de la revista: Medicina Año: 2020 Tipo del documento: Artículo País de afiliación: J.issn.1008-9292.2020.10.09