Evaluation of the diagnostic accuracy of a new point-of-care rapid test for SARS-CoV-2 virus detection.
J Transl Med
; 18(1): 488, 2020 12 21.
Artículo
en Inglés
| MEDLINE | ID: covidwho-992499
ABSTRACT
BACKGROUND:
The easy access to a quick diagnosis of coronavirus disease 2019 (COVID-19) is a key point to improve the management of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) and to contain its spread. Up to now, laboratory real-time PCR is the standard of care, but requires a fully equipped laboratory and significant infrastructure. Consequently, new diagnostic tools are required.METHODS:
In the present work, the diagnostic accuracy of the point-of-care rapid test "bKIT Virus Finder COVID-19" (Hyris Ltd) is evaluated by a retrospective and a prospective analysis on SARS CoV-2 samples previously assessed with an FDA "authorized for the emergency use-EUA" reference method. Descriptive statistics were used for the present study.RESULTS:
Results obtained with the Hyris Kit are the same as that of standard laboratory-based real time PCR methods for all the analyzed samples. In addition, the Hyris Kit provides the test results in less than 2 h, a significantly shorter time compared to the reference methods, without the need of a fully equipped laboratory.CONCLUSIONS:
To conclude, the Hyris kit represents a promising tool to improve the health surveillance and to increase the capacity of SARS-CoV-2 testing.Palabras clave
Texto completo:
Disponible
Colección:
Bases de datos internacionales
Base de datos:
MEDLINE
Asunto principal:
Sistemas de Atención de Punto
/
Prueba de Ácido Nucleico para COVID-19
/
COVID-19
Tipo de estudio:
Estudio de cohorte
/
Estudios diagnósticos
/
Estudio experimental
/
Estudio observacional
/
Estudio pronóstico
Límite:
Humanos
País/Región como asunto:
Europa
Idioma:
Inglés
Revista:
J Transl Med
Año:
2020
Tipo del documento:
Artículo
País de afiliación:
S12967-020-02651-y
Similares
MEDLINE
...
LILACS
LIS