Your browser doesn't support javascript.
loading
Montrer: 20 | 50 | 100
Résultats 1 - 20 de 26
Filtre
1.
Chinese Journal of Radiological Medicine and Protection ; (12): 145-149, 2020.
Article Dans Chinois | WPRIM | ID: wpr-799421

Résumé

Objective@#To study a method for verifying the doses to PTV and OAR as well as the 2D dose distribution arising from IMRT through using radiochromic films and TLDs.@*Methods@#Totally 7 medical electronic linear accelerators from Varian, Siemens and Elekta were selected. The polystyrene phantom provided by IAEA was conducted with CT scan. After irradiation with 6 MV X-rays, the TLDs and films were returned to the secondary standard dosimetry laboratory of China CDC for measurement and estimation.@*Results@#According to the IAEA requirements, the relative deviations between TLD-measured and TPS-planned values for PTV and OAR doses were both within ±7.0%. For PTV, the measured relative deviations for 5 accelerators were in the range of -4.0% to 3.4%, consistent with the IAEA requirements, whereas the values for the other 2 accelerators were in the range of -7.0% to 10.6%, not consistent with the requirements. For OAR, the values for 4 accelerators were in the range of -5.6% to 3.3%, consistent with the IAEA requirements, whereas the values for the other 3 accelerators were in the range of -20.8% to 11.5%, not meeting the requirements. As required by the IAEA, the 2D dose distribution 3 mm/3% pass rate should be higher than 90%. The measured values for 5 accelerators were in the range of 91.8% to 98.5%, consistent with the requirements, whereas the values measured for the other 2 were 45.0% and 77.0% respectively, not meeting the requirements.@*Conclusions@#It is feasible for using TLDs and radiochromic films to verify the doses to PTV and OAR and the 2D dose distribution in IMRT. This method should be applied to not only quality verification but also hospital internal audit to the extent possible.

2.
Chinese Journal of Radiological Medicine and Protection ; (12): 140-144, 2020.
Article Dans Chinois | WPRIM | ID: wpr-799420

Résumé

Objective@#To use TLDs and radiochromic films to verify the prescribed doses to both planned target volume (PTV) and organ at risk (OAR) and the 2D dose distribution in IMRT.@*Methods@#Eight accelerators of different models were selected in Henan province. The polystyrene phantom provided by IAEA was scanned using CT scanners and then the scanned images were transmitted to treatment planning system (TPS) for prescribing respectively the doses to PTV and OAR. IMRT was performed with phantom exposed to a 6 MV X-rays. The irradiated TLDs and films were delivered for measurement and estimation at Secondary Standard Dosimetry Laboratory at National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention.@*Results@#According to IAEA requirements, the relative deviations of the TLD-measured and TPS-planned values were within ±7.0% for the prescribed doses to PTV and OAR. The measured results for PTV have shown that the relative deviation of TLD-measured and TPS-planned values were within -0.3% to 6.9% for 8 accelerators, all consistent with the IAEA requirements. For OAR, the relative deviations of TLD-measured and TPS-planned were within -7.0% to 0.3% for 6 accelerators, consistent with the requirements, whereas those for other 2 accelerators were within -10.8% to -8.4%, not up to the requirements. IAEA required that, for 2D dose distribution, the pass rate of 3 mm/3% be ≥90%. The measured values for 7 accelerators were from 90.2% to 99.9%, consistent with the requirements, whereas that for another one was 70.0%, not meeting the requirement.@*Conclusions@#The method to verify, using radiochromic film and TLD, the prescribed doses to PTV and OAR and the pass rate of 2D dose distribution is simple and reliable. It is an important step to implement quality control for IMRT and can provide effective support for medical or third-party service institution to verify clinically prescribed dose.

3.
Chinese Journal of Radiological Medicine and Protection ; (12): 134-139, 2020.
Article Dans Chinois | WPRIM | ID: wpr-799419

Résumé

Objective@#To study a method to measure the doses to planned target volume (PTV) and organ at risk (OAR) and 2D dose distribution in IMRT by using TLD and radiochromic film for a verification purpose.@*Methods@#Totally 7 different types of medical linear accelerators were selected from seven hospitals in Hubei province. A polystyrene phantom provided by IAEA was CT scanned and then the scanned images were returned to the Treatment Planning System (TPS) for determining the prescribed doses to PTS and OAR and the corresponding MU. After the phantom was irradiated with 6 MV X-ray, the TLDs and films were transmitted to the secondary standard dosimetry laboratory of China CDC for measurement and estimation.@*Results@#The IAEA required the relative deviations between TLD-measured and TPS-planned doses to OAR and PTV be within ±7.0%. For PTV, the measured-to-planned deviation values for 7 accelerator were within -5.4% to 6.5%, all consistent with the IAEA requirements. For OAR, the values for 5 accelerators were within -2.2% to 6.7%, not consistent the requirements, whereas the values for the other 2 were -8.6% and 8.2% respectively, beyond the required values. The IAEA required that the 2D dose distribution 3 mm/3% pass rate be higher than 90%. The measured values for 7 accelerators were in the range of 90.3%-98.9%, all consistent with the requirements.@*Conclusions@#It is feasible scientifically and easy to operate in practice for using TLD and film to carry out dose verification in IMRT. It would be advisable to apply this method to quality verification in IMRT in medical institutions to the extent possible.

4.
Chinese Journal of Radiological Medicine and Protection ; (12): 129-133, 2020.
Article Dans Chinois | WPRIM | ID: wpr-799418

Résumé

Objective@#To validate the method for measuring the TPV and OAR doses and 2D dose distribution in IMRT through using TLD and radiochromic film.@*Methods@#Eight medical linear accelerators (Valian, Elekta, Siemens) were selected. The polystyrene phantom provided by IAEA was CT scanned and the image obtained was transferred to TPS for formulation of treatment plan, prescription of PTV and OAR doses and calculation of corresponding monitoring unit (MU), IMRT was performed on the phantom using 6 MV X-ray. Irradiated TLDs and films were measured and evaluated at the Secondary Standard Dosimetry Laboratory at the Radiation Safety Institute of Chinese Center for Disease Control and Prevention.@*Results@#According to IAEA requirement, the relative deviations between TLD-measured and TPS-planned doses were within ±7.0% for the prescribed PTV and OAR doses. As measured result, the PTV values for 8 accelerators were in the range of 0.6% to 5.9%, consistent with the IAEA requirements, whereas the OAT values for 8 accelerators were within -0.6% to 7.0%, consistent the requirements. As IAEA required, the 2D dose distribution passing rate of 3 mm/3% should be higher than 90%. The film-measured and TPS-planned values for 8 accelerators were within 90.2% to 100.0%, consistent with the requirements.@*Conclusions@#TLD and radiochromic film are feasible in validating the PTV and OAR doses and the 2D dose distribution pass rate in IMRT. This method can be widely used in quality audit and internal verification in IMRT in medical institutiions on a large scale.

5.
Chinese Journal of Radiological Medicine and Protection ; (12): 122-128, 2020.
Article Dans Chinois | WPRIM | ID: wpr-799417

Résumé

Objective@#To develop the methodology for using TLD and radiochromic film to measure the planned target volume (PTV) and organ at risk (OAR) doses and 2D dose distribution in IMRT, in order to provide technical guidance on the dose quality audit in IMRT at home.@*Methods@#China has participated in the research project launched by the international multi-radiotherapy centre (IMRC). IMRT polystyrene phantom provided by IAEA was scanned by CT scanner and then the scanned images were transmitted to TPS to outline prescribed dose to PTV and to OAR. The former was limited to 400 cGy while the latter limited to 200 cGy. IMRT was implemented with the phantom irradiated using 6 MV X-ray. The irradiated TLDs and films were sent to IAEA dosimerty laboratory for measurement and calculation. Jiangsu, Sichuan, Hubei and Henan provinces were selected to engage with this study for their variety of accelerators and highly skilled physicists. The procedures used were the same as in the IMRC and the irradiated TLDs and films were required to send to external audit group for measurement and calculation.@*Results@#According to IAEA requirement, the relative deviations of the TLD-measured and TPS planned doses are within ±7.0% for PTV and OAR. The China′s research results at the IMRC have shown that the relative deviation of TLD-measured and TPS-planned values for the upper and lower PTV were -0.2% and 0.8%, respectively, consistent with the IAEA requirement, and the values for upper and lower OAR were -0.6% and -1.0%, respectively, consistent with the requirement. As the results have shown in four provinces, the relative deviations of the TLD-measured and TPS-planned were within 0 to 10.6% for upper and lower PTV and -0.6% to 20.9% for upper and lower OAR. According to IAEA requirement, the passing rate should be greater than 90% for 3 mm /3% for 2D dose distribution. China′s result at the IMRC is 100%, being excellent. The four provinces′ results have shown that 2D dose distribution pass rate of 3 mm/3% was in the range of 45.0%-100.0%.@*Conclusions@#The uses of TLD in quality audit for PTV and OAR doses and the radiochromic film in 2D dose distribution pass rate in IMRT are characterized by scientific feasibility, strong operability, easy-to-mail and data realibility. They are can be applied to quality assurance and audit in medical institutions in the country to on a large scale.

6.
Chinese Journal of Radiological Medicine and Protection ; (12): 145-149, 2020.
Article Dans Chinois | WPRIM | ID: wpr-868416

Résumé

Objective To study a method for verifying the doses to PTV and OAR as well as the 2D dose distribution arising from IMRT through using radiochromic films and TLDs.Methods Totally 7 medical electronic linear accelerators from Varian,Siemens and Elekta were selected.The polystyrene phantom provided by IAEA was conducted with CT scan.After irradiation with 6 MV X-rays,the TLDs and films were returned to the secondary standard dosimetry laboratory of China CDC for measurement and estimation.Results According to the IAEA requirements,the relative deviations between TLD-measured and TPS-planned values for PTV and OAR doses were both within ±7.0%.For PTV,the measured relative deviations for 5 accelerators were in the range of-4.0% to 3.4%,consistent with the IAEA requirements,whereas the values for the other 2 accelerators were in the range of-7.0% to 10.6%,not consistent with the requirements.For OAR,the values for 4 accelerators were in the range of-5.6% to 3.3%,consistent with the IAEA requirements,whereas the values for the other 3 accelerators were in the range of-20.8% to 11.5%,not meeting the requirements.As required by the IAEA,the 2D dose distribution 3 mm/3% pass rate should be higher than 90%.The measured values for 5 accelerators were in the range of 91.8% to 98.5%,consistent with the requirements,whereas the values measured for the other 2 were 45.0% and 77.0% respectively,not meeting the requirements.Conclusions It is feasible for using TLDs and radiochromic films to verify the doses to PTV and OAR and the 2D dose distribution in IMRT.This method should be applied to not only quality verification but also hospital internal audit to the extent possible.

7.
Chinese Journal of Radiological Medicine and Protection ; (12): 140-144, 2020.
Article Dans Chinois | WPRIM | ID: wpr-868415

Résumé

Objective To use TLDs and radiochromic films to verify the prescribed doses to both planned target volume (PTV) and organ at risk (OAR) and the 2D dose distribution in IMRT.Methods Eight accelerators of different models were selected in Henan province.The polystyrene phantom provided by IAEA was scanned using CT scanners and then the scanned images were transmitted to treatment planning system (TPS) for prescribing respectively the doses to PTV and OAR.IMRT was performed with phantom exposed to a 6 MV X-rays.The irradiated TLDs and films were delivered for measurement and estimation at Secondary Standard Dosimetry Laboratory at National Institute for Radiological Protection,Chinese Center for Disease Control and Prevention.Results According to IAEA requirements,the relative deviations of the TLD-measured and TPS-planned values were within ±7.0% for the prescribed doses to PTV and OAR.The measured results for PTV have shown that the relative deviation of TLD-measured and TPS-planned values were within-0.3% to 6.9% for 8 accelerators,all consistent with the IAEA requirements.For OAR,the relative deviations of TLD-measured and TPS-planned were within-7.0% to 0.3% for 6 accelerators,consistent with the requirements,whereas those for other 2 accelerators were within-10.8% to-8.4%,not up to the requirements.IAEA required that,for 2D dose distribution,the pass rate of 3 mm/3% be ≥ 90%.The measured values for 7 accelerators were from 90.2% to 99.9%,consistent with the requirements,whereas that for another one was 70.0%,not meeting the requirement.Conclusions The method to verify,using radiochromic film and TLD,the prescribed doses to PTV and OAR and the pass rate of 2D dose distribution is simple and reliable.It is an important step to implement quality control for IMRT and can provide effective support for medical or third-party service institution to verify clinically prescribed dose.

8.
Chinese Journal of Radiological Medicine and Protection ; (12): 134-139, 2020.
Article Dans Chinois | WPRIM | ID: wpr-868414

Résumé

Objective To study a method to measure the doses to planned target volume (PTV) and organ at risk (OAR) and 2D dose distribution in IMRT by using TLD and radiochromic film for a verification purpose.Methods Totally 7 different types of medical linear accelerators were selected from seven hospitals in Hubei province.A polystyrene phantom provided by IAEA was CT scanned and then the scanned images were returned to the Treatment Planning System (TPS) for determining the prescribed doses to PTS and OAR and the corresponding MU.After the phantom was irradiated with 6 MV X-ray,the TLDs and films were transmitted to the secondary standard dosimetry laboratory of China CDC for measurement and estimation.Results The IAEA required the relative deviations between TLD-measured and TPS-planned doses to OAR and PTV be within ±7.0%.For PTV,the measured-to-planned deviation values for 7 accelerator were within-5.4% to 6.5%,all consistent with the IAEA requirements.For OAR,the values for 5 accelerators were within-2.2% to 6.7%,not consistent the requirements,whereas the values for the other 2 were-8.6% and 8.2% respectively,beyond the required values.The IAEA required that the 2D dose distribution 3 mm/3% pass rate be higher than 90%.The measured values for 7 accelerators were in the range of 90.3%-98.9%,all consistent with the requirements.Conclusions It is feasible scientifically and easy to operate in practice for using TLD and film to carry out dose verification in IMRT.It would be advisable to apply this method to quality verification in IMRT in medical institutions to the extent possible.

9.
Chinese Journal of Radiological Medicine and Protection ; (12): 129-133, 2020.
Article Dans Chinois | WPRIM | ID: wpr-868413

Résumé

Objective To validate the method for measuring the TPV and OAR doses and 2D dose distribution in IMRT through using TLD and radiochromic film.Methods Eight medical linear accelerators (Valian,Elekta,Siemens) were selected.The polystyrene phantom provided by IAEA was CT scanned and the image obtained was transferred to TPS for formulation of treatment plan,prescription of PTV and OAR doses and calculation of corresponding monitoring unit (MU),IMRT was performed on the phantom using 6 MV X-ray.Irradiated TLDs and films were measured and evaluated at the Secondary Standard Dosimetry Laboratory at the Radiation Safety Institute of Chinese Center for Disease Control and Prevention.Results According to IAEA requirement,the relative deviations between TLD-measured and TPS-planned doses were within ±7.0% for the prescribed PTV and OAR doses.As measured result,the PTV values for 8 accelerators were in the range of 0.6% to 5.9%,consistent with the IAEA requirements,whereas the OAT values for 8 accelerators were within-0.6% to 7.0%,consistent the requirements.As IAEA required,the 2D dose distribution passing rate of 3 mm/3% should be higher than 90%.The filmmeasured and TPS-planned values for 8 accelerators were within 90.2% to 100.0%,consistent with the requirements.Conclusions TLD and radiochromic film are feasible in validating the PTV and OAR doses and the 2D dose distribution pass rate in IMRT.This method can be widely used in quality audit and internal verification in IMRT in medical institutiions on a large scale.

10.
Chinese Journal of Radiological Medicine and Protection ; (12): 122-128, 2020.
Article Dans Chinois | WPRIM | ID: wpr-868412

Résumé

Objective To develop the methodology for using TLD and radiochromic film to measure the planned target volume (PTV) and organ at risk (OAR) doses and 2D dose distribution in IMRT,in order to provide technical guidance on the dose quality audit in IMRT at home.Methods China has participated in the research project launched by the international multi-radiotherapy centre (IMRC).IMRT polystyrene phantom provided by IAEA was scanned by CT scanner and then the scanned images were transmitted to TPS to outline prescribed dose to PTV and to OAR.The former was limited to 400 cGy while the latter limited to 200 cGy.IMRT was implemented with the phantom irradiated using 6 MV X-ray.The irradiated TLDs and films were sent to IAEA dosimerty laboratory for measurement and calculation.Jiangsu,Sichuan,Hubei and Henan provinces were selected to engage with this study for their variety of accelerators and highly skilled physicists.The procedures used were the same as in the IMRC and the irradiated TLDs and films were required to send to external audit group for measurement and calculation.Results According to IAEA requirement,the relative deviations of the TLD-measured and TPS planned doses are within ±7.0% for PTV and OAR.The China's research results at the IMRC have shown that the relative deviation of TLD-measured and TPS-planned values for the upper and lower PTV were-0.2% and 0.8%,respectively,consistent with the IAEA requirement,and the values for upper and lower OAR were -0.6% and-1.0%,respectively,consistent with the requirement.As the results have shown in four provinces,the relative deviations of the TLD-measured and TPS-planned were within 0 to 10.6% for upper and lower PTV and-0.6% to 20.9% for upper and lower OAR.According to IAEA requirement,the passing rate should be greater than 90% for 3 mm /3% for 2D dose distribution.China's result at the IMRC is 100%,being excellent.The four provinces' results have shown that 2D dose distribution pass rate of 3 mm/3% was in the range of 45.0%-100.0%.Conclusions The uses of TLD in quality audit for PTV and OAR doses and the radiochromic film in 2D dose distribution pass rate in IMRT are characterized by scientific feasibility,strong operability,easy-to-mail and data realibility.They are can be applied to quality assurance and audit in medical institutions in the country to on a large scale.

11.
Chinese Journal of Radiological Medicine and Protection ; (12): 121-127, 2019.
Article Dans Chinois | WPRIM | ID: wpr-734326

Résumé

Objective To develop measurement methodology using film for the positioning accuracy of MLC leaves in IMRT.Methods The solid water phantom of 30 cm x 30 cm was scanned and the scanned images were transferred to TPS for treatment plan formulation.The five MLC strip picket fence pattern was formed by MLC leaves,each 3.0 cm long × 6.0 mm wide.The separation between strip and strip is 3.0 cm.SAD is 100 cm at dmax for 6 MV X-ray,with 250 MU per MLC strip.EBT2 radiochomic film was put on the phantom for delivery of IMRT,for each MLC strip.The present study focused on 30 accelerators of Varian,Elekta and Siemens designs at 27 hospitals with highly skilled physicists all over Jiangsu,Sichuan,Hubei and Henan provinces.The study was conducted in the same way as used in international multi-radiotherapy center (IMRC).The irradiated films were sent respectively to IAEA dosimetry laboratory and external audit group (EGA) of China for measurement,analysis and calculation.Results According to IAEA requirements,the differenc of film-measured and TPS-planned of MLC leaf position for each strip should be within ±0.5 mm.China had participated in the research of IMRC,with the result of 0.3,0.2,0.0,-0.1,and-0.2 mm,respectively.For 30 accelerators in four provinces involved in the study,the IAEA's verification results of MLC leaf position were within 0.6-1.0 mm for 5 accelerators and within ± 0.5 mm for other 25 ones.Whereas the verification results of EAG were within 0.6-1.0 mm for 6 accelerators and within ±0.5 mm for other 24 ones.According to IAEA requirements,the film-measured MLC leaf position deviation for each pair of leaves and average all pairs of leaves should be within ±0.5 mm.China had participated in IMRC's research,with the measured result being 0.04 mm.The verification result of EGA for 30 accelerators showed the measured MLC leaf position deviations were all <0.3 mm per strip,consistent with IAEA requirements.The IAEA's result showed the measured deviations of MLC leaf position for 29 accelerators were within ±0.5 mm,with only other one being-0.7 mm not consistent with the IAEA requirements.As required by IAEA,the difference of film-measured difference of MLC opening width should be within ±0.75 mm between each pair and average all pairs of leaves.China's result in research of IMRC showed the difference of minimum width to mean width was-0.2 mm whereas the difference of maximum width to mean width was 0.4 mm.For 30 accelerators involved in IAEA'verification study,the measured result shown that the difference between maximum and average of filmmeasured of MLC leaf width,and between minmum and average,were within ± 0.75 mm for 24 accelerators,in line with the IAEA requirements.For other 6 ones,the values were beyond ±0.75 mm,not in line with the IAEA requirements.For the verification result of EAG,the difference between maximum and average widths and between minmum and average widths for 25 accelerators were within ±0.75 mm as required by IAEA,whereas for other 5 ones the value were beyond ±0.75 mm,not consistent with IAEA requirements.The standard deviation of film-measured MLC opening width between each pair and average all pairs should be within 0.3 mm as required by IAEA.China's IMRC result was 0.12 mm.The verification result of IAEA shown the standard deviation of MLC opening position were <0.3 mm for 26 accelerators and > 0.3 mm for other 4 accelerators.EAG verification result were the same as IAEA result.Conclusions The method using radiochromic film for measuring accuracy of MLC leaf position is convenient and practicable as a quality audit.It is suitable for quality verification in medical institutions owing to easy to post and repeated measurements.

12.
Chinese Journal of Radiological Medicine and Protection ; (12): 303-308, 2019.
Article Dans Chinois | WPRIM | ID: wpr-745257

Résumé

Objective To estimate the hand dose of occupational staff in five procedures of interventional radiology.Methods The thermoluminescence ring dosimeter(TLD,LiF:Mg,Ti)calibrated in dose equivalent Hp(0.07) was used to monitor the radiation dose to the both hands in five procedures of interventional radiology in four hospitals.Meanwhile,the tube voltage and current and the fluoroscopy time,accumulated doses,DAPs (Dosed-Area Product) and photographic frames were also recorded for statistical analysis(SPSS 18.0).Results In this study,five interventional procedures were monitored in 119 cases.The doses to the left and right hands of occupational personnel in the five interventional procedures were analyzed,and the difference was statistically significant (t =1.99,P< 0.05).The differences in the dose to the left hand and the right hand of the first operator with different interventional procedures were statistically significant (F =455.83,116.45,P<0.01).Tube voltage,current,fluoroscopy time,and photographic frames in influence factor analysis were statistically significant (r =0.570,0.712,0.564,0.711,P< 0.05),indicating that four factors might affect the operator's hand doses.In addition,increasing these four factors directly led to increase in the operator's hand doses.The variables with statistical significance in the above single factor analysis were introduced into the multiple linear regression equation and the stepwise regression method was used to fit the equation.The fitting equation was y=225.763+ 1.862x1-98.125x2 (F=22.726,P<0.05).Where,x1 was the fluoroscopy time and x2 was the photographic frames which showed that they were the main factors affecting the hand dose.Conclusions The dose to the hands of the primary operator was the highest,followed by the second operator,the assistant or nurses.The order of average doses to the hands of the primary operator was PM>RFA>CA> PTCA+PCI > ITCA in the five procedures.The annual equivalent dose to primary operator's hands may exceed the dose limit for hands in practicing large amount of the PM procedures.

13.
Chinese Journal of Emergency Medicine ; (12): 1292-1295, 2019.
Article Dans Chinois | WPRIM | ID: wpr-789213

Résumé

Objective To compare the clinical effect of conventional polyvinyl chloride (PVC) endotracheal tube with silicone wire reinforced endotracheal tube to prevent ventilator-associated pneumonia (VAP) in mechanical ventilation patients.Methods A total of 240 mechanical ventilation patients in the ICU were enrolled in this study,which were divided into two groups,PVC endotracheal intubation group (PVC group,n=1 13) and wire reinforced endotracheal intubation group (WR group,n=127).Gender,age,APACHE Ⅱ score,one-time success rate of intubation,intubation time,the rate of changing endotracheal,ventilation time,rate of tracheotomy and the incidence of VAP were compare between the two groups.Predictive factors for VAP were identified by the univariate and multivariate analyses in step-wise logistic regression model.Results The rate of changing endotracheal tube in the PVC group was lower than that in the WR group (x2=5.785,P=0.016);the mechanical ventilation time in the PVC group was shorter than that in the WR group (t=2.180,P=0.018);and compared with the WR group,the PVC group had significantly lower incidence of VAP (x2=6.215,P=0.012).The univariate analysis showed that the selection ofPVC endotracheal tube,APACHE Ⅱ score and mechanical ventilation time were the significant risk factors for VAP (P < 0.05).Multivariate analysis showed that the different selection of endotracheal tube and mechanical ventilation time were independent influencing factors ofVAP (P < 0.05).Conclusions PVC tracheal tube can effectively reduce the incidence of VAP in patients with mechanical ventilation.

14.
Chinese Journal of Radiological Medicine and Protection ; (12): 368-373, 2018.
Article Dans Chinois | WPRIM | ID: wpr-708071

Résumé

Objective To measure the doses to eye lens and hands of workers,using thermo luminescent dosimeter (TLD) and optically stimulated luminescence dosimeter (OSLD).Methods TLDs in the same batch were annealed,packed and stuck to the flat abdomen of Alderson-Phantom at a distance of about 15 cm from 125I seed source,while irradiated at different doses:1.0,1.5,3.0,5.0,10.0,12.0,20.0,25.0,30.0,50.0 and 60.0 μ Gy.And then TLDs were measured by dosimeters to establish a dose calibration curve.By implanting seed source into the selected lung for 14 cases,belly for 10 cases,pelvic for 5 case and neck for 6 cases while placing calibrated TLDs on the left,middle and right above eyes,left and right hands of the workers to obtain the location-specific kerma values.Finally,the conversion factors Hp (3) and Hp (0.07) were used to calculate the values of dose equivalent to eye lens and hands.Additionally,OSLDs were used to measure the doses to workers in the same way.Results The TLD-measured eye lens dses to the operator and his assistant were 0.8 and 1.6 mSv in lungs,1.3 and 1.2 mSv in bellies,0.9 and 0.6 mSv in pelves,0.3 mSv in necks,respectively.Meanwhile,hand doses to the operator and his assistant were 1.4 and 2.1 mSv in lungs,1.2 and 1.0 mSv in bellies,0.5 and 0.9 mSv in pelves,0.1 mSv in necks,respectively.The maximum doses to eye lens and hands were 1.2 and 1.0 mSv,respectively in a single treatment.OSLD-measured dose equivalents from lung therapy were 0.2 and 0.1 mSv for eye lens of the operator and his assistant and 0.4 and 0.6 mSv for hands.For belly therapy,the accumulated dose equivalent to hands of the operator was 0.1 mSv while those for other types of therapy were 0 mSv.Conclusions TLDs have the capability to measure not only accumulated dose but also dose equivalent from a single therapy According to ICRP 118 publication and as estimated in the present study,the number of therapy should be not more than 17 every year.OSLDs only give the accumulated dose,the accuracy of which needs to be studied in low-dose measurement.

15.
Chinese Journal of Radiological Medicine and Protection ; (12): 115-120, 2018.
Article Dans Chinois | WPRIM | ID: wpr-708025

Résumé

Objective To develop the methodology for using TLDs and films to measure absorbed dose and 2D dose distribution produced by the multi-leaf collimator (MLC) in intensity modulated radiotherapy (IMRT),in order to provide the guidance on dose quality audit in IMRT.Methods A total of 30 different-typed accelerators were selected from 27 hospitals in Jiangsu,Sichuan,Hubei and Henan provinces,including 17 Varian accelerators,10 Elektas and 3 Simens.The same batch of films and TLDs were put in a 2 cm-thick solid plate for fixation and then loaded in a 15 cm × 15 cm × 15 cm polysyrene solid phantom supplied by International Atomic Energy Agency(IAEA) in terms of 90 cm SSD,19 cm depth,10 cm × 10 cm field at different doses.The standard dose curves wcrc established for film and TLD,respectively.The irradiated film was measured and then sent to the External Audit Group (EAG) in China.The TLD-and film-absorbed doses were compared with TPS-calculated doses.The 2D dose distribution on the IRMT MLC field was measured using films.The homogeneous phanton of 30 cn × 30 cm was scanned by CT and the image was transferred to the TPS.The IMRT was implemented with 6 Gy fractionated irradiation by placing a 25 cm × 25 cm film on the phantom surface at 95 cm SSD and at 5 cm depth.The irradiated film was sent to the IAEA dosimetry laboratory for measurement and calculation.2D dose distribution verification was conducted in thc same way consistent with the procedure of international multi-radiotherapy center.The 3 mm/3% passing rate was calculated for 2D dose distribution and compared with the film-measured and TPS calculated result.Results IAEA requires the relative deviation of TLD and film measured absorbed dose are with in ± 5%.The relative deviation of TLD-and filmmeasured to TPS-calculated absorbed dose was within the range of ±0.7%-± 8.5% and within ±0.3% ±7.8% in Jiangsu,Sichuan,Hubei and Henan provinces,respectively.IAEA requires the 3 mm/3% passing rate of film-measured 2D distribution to be 90%.The result of the present study were up to 94.0%.The verification result of 2D dose distribution were within 70.0%-99.9% in Sichuan,Jiangsu,Hubei and Henan provinces.Conclusions The adsorbed dose and 2D distribution can be audited using TLDs and films for MLC in IRMT.The method is scientific and applicable,economical and convenient for development of dose quality audit for a wide range of IRMT.

16.
Academic Journal of Second Military Medical University ; (12): 961-966, 2017.
Article Dans Chinois | WPRIM | ID: wpr-838464

Résumé

Objective To explore the causes of nocturnal hemodialysis (NHD) patient dropout and the risk factors for dropout. Methods We collected the clinical data of patients receiving NHD for more than 3 months, of whom 47 patients dropped out and 64 kept receiving NHD from Feb. 2009 to Nov. 2016 in Changzheng Hospital of Second Military Medical University. We investigated the general conditions; and we compared the differences of the blood parameters between the two groups when the patients received NHD for the first time and for the last time, including hemoglobin, platelet, albumin, ferritin, serum calcium, serum phosphorus and parathyroid hormone. We also analyzed the risk factors for NHD dropout or for death using Cox regression analysis model. Results Among 111 patients, 47 patients had withdrawn from NHD, with their average time for NHD being (31.55±20.30) months, and the causes for dropout included death, transferring to other hospitals, turning to conventional hemodialysis (CHD), renal transplantation and others. Univariate Cox regression analysis showed that hypertensive nephropathy (P=0.007, HR=2.913, 95%CI: 1.348-6.293) and diabetic nephropathy (P=0.047, HR=2.401, 95%CI: 1.014-5.685) were risk factors for NHD patient dropout, while chronic nephritis syndrome (P<0.001, HR=0.095, 95%CI 0.046-0.195) was a protective factor; blood albumin (P=0.007, HR=0.904, 95%CI:0.840-0.973) and age (P=0.027, HR=1.052, 95%CI:1.006-1.101) were risk factors for NHD patient dropout. Multivariate Cox regression analysis showed that albumin level (P=0.007, HR=0.911, 95%CI: 0.848-0.991) was an independent risk factor for death in NHD patients. Conclusion Hypertensive nephropathy and diabetic nephropathy are the risk factors for NHD patients dropout, while chronic nephritis syndrome was a protective factor. Low serum albumin level is an independent risk factor for death in NHD patients.

17.
China Journal of Chinese Materia Medica ; (24): 719-724, 2017.
Article Dans Chinois | WPRIM | ID: wpr-275472

Résumé

Five main flavonoids of Hebei Xiangju were studied using the Density Functional Theory (DFT) B3LYP method with 6-311 G (d) basis set.Their activities were analyzed based on molecular structure,bond dissociation energy (BDE),natural orbital charge distribution (NBO),bond order and the energy gap between HOMO and LUMO. The results showed that the existing of intra molecular hydrogen bond in B ring can improve the antioxidant activity of the flavonoids, at the same time, the hydroxyl groups on the glycosides do not have the activity of eliminating free radicals, but decrease the total molecular antioxidant activity. As a result, the antioxidant ability order of the five flavonoids compounds is luteolin< luteolin-7-O-glucoside< apigenin < acacetin < acacetin-7-O-glucose, which is agreement with the experimental conclusion reported in literature. The results showed that the DFT method can provide theoretical guidance for the selection of natural flavonoid antioxidants.

18.
Academic Journal of Second Military Medical University ; (12): 961-966, 2017.
Article Dans Chinois | WPRIM | ID: wpr-607059

Résumé

Objective To explore the causes of nocturnal hemodialysis (NHD) patient dropout and the risk factors for dropout.Methods We collected the clinical data of patients receiving NHD for more than 3 months,of whom 47 patients dropped out and 64 kept receiving NHD from Feb.2009 to Nov.2016 in Changzheng Hospital of Second Military Medical University.We investigated the general conditions;and we compared the differences of the blood parameters between the two groups when the patients received NHD for the first time and for the last time,including hemoglobin,platelet,albumin,ferritin,serum calcium,serum phosphorus and parathyroid hormone.We also analyzed the risk factors for NHD dropout or for death using Cox regression analysis model.Results Among 111 patients,47 patients had withdrawn from NHD,with their average time for NHD being (31.55±20.30) months,and the causes for dropout included death,transferring to other hospitals,turning to conventional hemodialysis (CHD),renal transplantation and others.Univariate Cox regression analysis showed that hypertensive nephropathy (P=0.007,HR=2.913,95%CI:1.348-6.293) and diabetic nephropathy (P=0.047,HR=2.401,95%CI:1.014-5.685) were risk factors for NHD patient dropout,while chronic nephritis syndrome (P<0.001,HR=0.095,95%CI 0.046-0.195) was a protective factor;blood albumin (P=0.007,HR=0.904,95%CI:0.840-0.973) and age (P=0.027,HR=1.052,95%CI:1.006-1.101) were risk factors for NHD patient dropout.Multivariate Cox regression analysis showed that albumin level (P=0.007,HR=0.911,95%CI:0.848-0.991) was an independent risk factor for death in NHD patients.Conclusion Hypertensive nephropathy and diabetic nephropathy are the risk factors for NHD patients dropout,while chronic nephritis syndrome was a protective factor.Low serum albumin level is an independent risk faetor for death in NHD patients.

19.
Chinese Journal of Pathophysiology ; (12): 2252-2258, 2017.
Article Dans Chinois | WPRIM | ID: wpr-663609

Résumé

AIM: To investigate whether Toll-like receptor 4 ( TLR4 ) and Nod-like receptor protein 3 (NLRP3) inflammasome were involved in contrast medium (CM)-induced inflammation and injury in renal tubular epithe-lial cells.METHODS: Iopromide was used to injure NRK-52E cells in the study.The cell viability was measured by CCK-8 assay.The protein levels of TLR4, NLRP3, apoptosis-associated speckle-like protein (ASC), caspase-1 and cleaved caspase-3 were determined by Western blot .The releases of interleukin ( IL )-1βand IL-18 were detected by ELISA .The apoptotic rate was evaluated by Hoechst staining , and mitochondrial membrane potential ( MMP) was analyzed by JC-1 staining.siRNA was transfected into the NRK-52E cells to silence NLRP3 expression.RESULTS:CM decreased the viability of NRK-52E cells (P<0.05).CM also elevated the protein levels of cleaved caspase-3, TLR4, NLRP3, IL-1βand IL-18 (P<0.05).Silencing NLRP3 attenuated CM-induced releases of inflammatory cytokines .Moreover, treat-ment with TLR4 inhibitor TAK-242 or knockdown of NLRP3 by siRNA transfection both attenuated cell apoptosis and loss of MMP caused by CM .CONCLUSION:TLR4/NLRP3 inflammasome takes part in the pathogenesis of CM-induced acute kidney injury , and mediates CM-induced injury and inflammation in renal tubular epithelial cells .

20.
Chinese Journal of Radiological Medicine and Protection ; (12): 775-779, 2015.
Article Dans Chinois | WPRIM | ID: wpr-480996

Résumé

Objective To develop the methods for using 0.015 cc pinpoint chambers, 0.007 cc miniature chambers and diode detector to measure Multi-leaf collimator (MLC) small field in IMRT.Methods MAX4000 and Unidos electrometers were connected with different types of small chambers and diode detectors.MLC shaped fields of10 cm×10 cm, 6 cm×6 cm, 4 cm×4 cm, 3 cm×3 cm, 2 cm× 2 cm were defined at 100 cm SSD.The field sizes for the Varian accelerator were defined by the tertiary MLC, while the secondary jaws were kept at 10 cm × 10 cm field, with the monitor units of 250 MU.Each field was measured three times to obtain the average value.The readings of all small fields were normalized to 10 cm × 10 cm field values for comparison of measured and published output factors.Results The relative deviations of the MLC small field output factors from the published outputs are 1.0% , 1.7% , 1.5% and 2.4%, respectively, for Unidos electrometer connected with 0.015 cc pinpoint chamber;0.2%, 0.8%, 0.8% and 1.4%, respectively, for Unidos electrometer connected with 0.007 cc miniature chamber;and 0.1%, 0.5%, 0.5% and 0.9%, respectively, for MAX4000 electrometer connected with 0.007 cc miniature chamber.Conclusions The 0.015 cc chamber-measured MLC output factors for 3 cm × 3 cm and 2 cm × 2 cm fields are excellent.As required by IAEA, the relative deviations of the measured output factor from the published output factor are within ± 2% for 2 cm × 2 cm fields and ± 3% for larger fields.The results measured using 0.007 cc chamber are better than those measured using 0.015 cc chamber.The measured results using the diode detector, normalized to the 10 cm × 10 cm field, are consistent with the minimum requirements and excellent when being normalized to the 4 cm × 4 cm field.For dosimetric consideration, MLC small field output factor should be measured using small chamber and diode detector.The method is accurate and reliable, therefore, all measured output factors for MLC small fields should be input into radiation treatment plan system.

SÉLECTION CITATIONS
Détails de la recherche